GMK STD CEMENTED FEMUR SIZE 5 LEFT
Report
- Report Number
- 3005180920-2015-00107
- Event Type
- Injury
- Date Received
- May 29, 2015
- Report Date
- September 4, 2015
- Manufacturer
- MEDACTA INTERNATIONAL, SA
- Product Code
- JWH
- PMA / PMN Number
- K090988
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
ON 05 AUGUST 2015 THE RETRIEVED ITEMS WERE INSPECTED BY THE (B)(4) WITH THE FOLLOWING COMMENT: FROM VISUAL INSPECTION OF BOTH THE TIBIAL BASEPLATE AND THE FEMUR EXPLANTED NO ANOMALIES WERE NOTED. IT WAS NOTED THAT NO RESIDUAL CEMENT WAS PRESENT IN THE CEMENTATION POCKETS. IN ONLY FEW PARTS OF THESE SURFACES SOME SIGNS WERE NOTED, PROBABLY DUE TO RESIDUAL CEMENT. THE REMAINING PART OF THE SURFACE LOOKS "CLEAN". IT IS POSSIBLE THAT IN THIS PORTION THE CEMENT DIDN'T ADHERE CORRECTLY, OR MAYBE IT WAS COMPLETELY REMOVED DURING THE WASHING PROCESS AFTER EXPLANTATION. AS DESCRIBED IN THE EVENT DESCRIPTION, CEMENT DID NOT BOND TO THE IMPLANT, NORMALLY RELATED TO THE SURGICAL TECHNIQUE. DUE TO THE LACK OF INFORMATION, IT IS NOT POSSIBLE TO IDENTIFY POSSIBLE ROOT CAUSES. ON 06 AUGUST 2015 IT WAS PREPARED A FINAL REPORT WITH THE INFORMATION COLLECTED DURING THE INVESTIGATION, ALREADY REPORTED IN THE INITIAL REPORT AND HERE ABOVE. ON 11 AUGUST 2015 THE REPORT WAS SENT TO THE INITIAL REPORTER AND THE CASE WAS CLOSED.
BATCH REVIEW PERFORMED ON (B)(4) 2015. LOT 112982: (B)(4) FEMURS MANUFACTURED AND RELEASED ON 08/10/2011. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED ISSUE. LOT 114634: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 02/08/2012. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED ISSUE. LOT 114250: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 01/12/2012. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED ISSUE. ON 05/12/2015 IT WAS CONFIRMED THAT NO X-RAY IS AVAILABLE. ON 05/15/2015 THE MEDICAL AFFAIRS MADE THE FOLLOWING ANALYSIS: IN THIS CASE, CEMENT DID NOT BOND TO THE IMPLANT. CEMENT IS NOT A GLUE, AND NO CHEMICAL INTERACTION SHOULD BE EXPECTED BETWEEN CEMENT AND PROSTHESIS. FOR THIS REASON, I WOULD NOT CONSIDERED ANY PARTICULAR ACTION TO BE TAKEN ON THE IMPLANT. THERE MAY BE SEVERAL REASONS WHY CEMENT DOES NOT BOND TO THE IMPLANT, AND NORMALLY THEY ARE RELATED WITH OPERATIVE TECHNIQUE. THEY INCLUDE TEMPERATURE ISSUES (E.G. COLD IMPLANTS OR UNUSUALLY WARM OP TEMPERATURE), PRESENCE OF LIQUIDS, TIMING ISSUES WHEN APPLYING CEMENT, AND SEVERAL OTHERS.
IMP REF # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 349900 | GMK STD CEMENTED FEMUR SIZE 5 LEFT | KNEE STD CEMENTED FEMUR | JWH | MEDACTA INTERNATIONAL, SA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |