FDA Adverse Event Injury Summary report: N

TRIATHLON PRIM CEM FXD BPLT #2

MDR report key: 4255570 · Received November 17, 2014

Report

Report Number
0002249697-2014-04290
Event Type
Injury
Date Received
November 17, 2014
Date of Event
June 25, 2014
Report Date
October 22, 2014
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
JWH
PMA / PMN Number
K031729
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(6). AN EVENT REGARDING TIBIAL BASEPLATE SUBSIDENCE INVOLVING A TRIATHLON BASEPLATE WAS REPORTED. THE EVENT WAS NOT CONFIRMED. METHOD & RESULTS: -DEVICE EVALUATION AND RESULTS: NOT PERFORMED AS NO ITEMS WERE RETURNED. -MEDICAL RECORDS RECEIVED AND EVALUATION: NOT PERFORMED AS NO MEDICAL RECORDS WERE PROVIDED. -DEVICE HISTORY REVIEW: THE DEVICE WAS MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. -COMPLAINT HISTORY REVIEW: THERE HAVE BEEN NO OTHER SIMILAR EVENTS FOR THE LOT REFERENCED. CONCLUSIONS: THE REPORTED EVENT COULD NOT BE CONFIRMED WITH THE INFORMATION THAT WAS PROVIDED. A ROOT CAUSE CANNOT BE DETERMINED WITHOUT MEDICAL RECORDS, X-RAYS, OPERATIVE NOTES AND FOLLOW-UP NOTES.

Additional Manufacturer Narrative · 1

IT WAS NOTED THAT THE DEVICE IS NOT AVAILABLE FOR EVALUATION BECAUSE IT WAS DISCARDED BY THE HOSPITAL. ADDITIONAL INFORMATION HAS BEEN REQUESTED AND IF RECEIVED, WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT UPON COMPLETION OF THE INVESTIGATION. NOT AVAILABLE.

Description of Event or Problem · 1

NTX REGISTRY, STUDY PATIENT (B)(6) HAD LEFT KNEE REPLACEMENT ON (B)(6) 2012 WITHIN THE STUDY. COMPLAINTS OF PAIN, INCREASED VALGUS AND DECREASED WALKING DISTANCE SINCE (B)(6) 2014. TIBIAL COMPONENT SUBSIDENCE WAS DIAGNOSED, EXPLANTATION SURGERY PERFORMED ON (B)(6) 2014.

Description of Event or Problem · 1

NTX REGISTRY, STUDY PATIENT (B)(6) HAD LEFT KNEE REPLACEMENT ON (B)(6) 2012 WITHIN THE STUDY. COMPLAINTS OF PAIN, INCREASED VALGUS AND DECREASED WALKING DISTANCE SINCE (B)(6) 2014. TIBIAL COMPONENT SUBSIDENCE WAS DIAGNOSED, EXPLANTATION SURGERY PERFORMED ON (B)(6) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
742440 TRIATHLON PRIM CEM FXD BPLT #2 IMPLANT JWH STRYKER ORTHOPAEDICS-MAHWAH EADMT

Patients

Seq Age Sex Outcome Treatment
1 80 YR Hospitalization| R