KNEE NAIL GOLD 10X38 RETRO/TIB
Report
- Report Number
- 1020279-2014-00695
- Event Type
- Injury
- Date Received
- November 14, 2014
- Date of Event
- October 22, 2014
- Report Date
- October 23, 2014
- Manufacturer
- SMITH & NEPHEW, INC.
- Product Code
- JWH
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
THE ASSOCIATED DEVICES WERE RETURNED AND EVALUATED. A REVIEW OF THE MANUFACTURING RECORDS DID NOT REVEAL ANY MANUFACTURING OR MATERIAL ABNORMALITIES. THE AS-RECEIVED KNEE NAIL WAS EXAMINED VISUALLY, STEREOMICROSCOPICALLY, AND WITH A SCANNING ELECTRON MICROSCOPE (SEM). AN ELEMENTAL ANALYSIS WAS PERFORMED ON THE NAIL FRACTURE SURFACE USING ENERGY DISPERSIVE X-RAY ANALYSIS (EDXA) TECHNIQUE IN CONJUNCTION WITH THE SEM. MACROSCOPIC EXAMINATION OF THE AS-RECEIVED NAIL SHOWED A FRACTURE AT THE HOLE IN THE END THE NAIL. MICROSCOPIC EXAMINATION SHOWED BURNISHED AREAS, EQUIAXED DIMPLES (MICRO-VOID COALESCENCE, OR MVC), AND FATIGUE STRIATIONS. BURNISHING IS CAUSED BY TWO SURFACES RUBBING TOGETHER POST-FRACTURE. MVC IS INDICATIVE OF A DUCTILE OVERLOAD. BEACH MARKS ARE A MACROSCOPIC PHENOMENA COMMONLY ASSOCIATED WITH FATIGUE LOADING. THESE FEATURES ARE TYPICAL OF METAL FATIGUE CRACKING. FATIGUE CRACKING IS CAUSED BY THE IMPLANT BEARING CYCLIC (I.E. REPEATED) STRESSES IN EXCESS OF THE MATERIAL ENDURANCE LIMIT FOR AN EXTENDED PERIOD OF TIME. EDXA SPECTRA WERE COLLECTED FROM THE FRACTURE SURFACE OF THE NAIL TO DETERMINE THE COMPOSITION OF THE NAIL AND SHOWED NO ABNORMALITIES. FROM THE ANALYSES CONDUCTED DURING THIS INVESTIGATION, IT WAS CONCLUDED THAT THE KNEE NAIL FRACTURED BY THE INITIATION AND SUBSEQUENT PROPAGATION OF FATIGUE CRACKING. THE FATIGUE CRACKING EVENTUALLY PROPAGATED TO AN EXTENT THAT THE REMAINING CROSS-SECTIONAL AREA OF THE NAIL COULD NOT BEAR THE IMPOSED PATIENT LOADING, LEADING TO OVERLOAD FRACTURE. FATIGUE CRACKING IS CAUSED BY THE NAIL BEARING CYCLIC (I.E. REPEATED) STRESSES IN EXCESS OF THE MATERIAL ENDURANCE LIMIT FOR AN EXTENDED PERIOD OF TIME. NO MATERIALS OR MANUFACTURING DEVIATIONS WERE FOUND IN THIS INVESTIGATION.
.
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO IMPLANT FRACTURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 738223 | KNEE NAIL GOLD 10X38 RETRO/TIB | KNEE IMPLANT | JWH | SMITH & NEPHEW, INC. | 07LM08354 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 27 YR | Required Intervention |