FDA Adverse Event Injury Summary report: N

DURACON CRUC NON BEAD B/P XL1

MDR report key: 4247853 · Received November 13, 2014

Report

Report Number
0002249697-2014-04264
Event Type
Injury
Date Received
November 13, 2014
Date of Event
October 20, 2014
Report Date
October 20, 2014
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
JWH
PMA / PMN Number
K926228
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PATIENT IS (B)(6) IN HEIGHT. AN EVENT REGARDING FRACTURE INVOLVING A DURACON BASEPLATE WAS REPORTED. THE EVENT WAS CONFIRMED. METHOD & RESULTS: DEVICE EVALUATION AND RESULTS: A PHOTOGRAPH OF THE BASEPLATE COMPONENT WAS PROVIDED. THE COMPONENT WAS FRACTURED THROUGH ONE CONDYLE. MEDICAL RECORDS RECEIVED AND EVALUATION: INSUFFICIENT MEDICAL RECORDS WERE RECEIVED FOR A MEDICAL REVIEW TO BE PERFORMED. DEVICE HISTORY REVIEW: NOT PERFORMED AS THE LOT ID IS UNKNOWN. COMPLAINT HISTORY REVIEW: NOT PERFORMED AS THE LOT ID IS UNKNOWN. CONCLUSIONS: THE PHOTOGRAPHS THAT WERE PROVIDED CONFIRM THAT THE DURACON BASEPLATE WAS FRACTURED. INSUFFICIENT INFORMATION WAS PROVIDED TO PERFORM A MEDICAL REVIEW OF THIS CASE. FURTHER INFORMATION, SUCH AS MEDICAL RECORDS AND DEVICE EVALUATION ARE NEEDED TO DETERMINE THE ROOT CAUSE.

Additional Manufacturer Narrative · 1

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT WAS REVISED ON A RIGHT KNEE DUE TO FAILED KNEE. WHEN SURGEON OPENED UP THE PATIENT, FOUND THAT THE BASEPLATE WAS BROKEN. BASEPLATE, FEMUR AND INSERT WERE REMOVED AND REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT WAS REVISED ON A RIGHT KNEE DUE TO FAILED KNEE. WHEN SURGEON OPENED UP THE PATIENT, FOUND THAT THE BASEPLATE WAS BROKEN. BASEPLATE, FEMUR AND INSERT WERE REMOVED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
734394 DURACON CRUC NON BEAD B/P XL1 IMPLANT JWH STRYKER ORTHOPAEDICS-MAHWAH UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 59 YR Hospitalization| O| R