PFC SIGMA MOD TIB TRAY CEM SZ2
Report
- Report Number
- 1818910-2014-31474
- Event Type
- Injury
- Date Received
- November 10, 2014
- Date of Event
- October 20, 2014
- Report Date
- October 20, 2014
- Manufacturer
- DEPUY IRELAND 9616671
- Product Code
- JWH
- PMA / PMN Number
- PK884796
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4).
ADDITIONAL NARRATIVE: NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE COMPLAINT DATABASE SEARCH FOUND NO OTHER REPORTED INCIDENT(S) AGAINST THE PROVIDED PRODUCT/LOT COMBINATION(S) SINCE RELEASE FOR DISTRIBUTION. BASED ON PREVIOUS INVESTIGATIONS THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY..
ADDITIONAL NARRATIVE: THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATION. THE INITIAL REPORTING STATED NO ADDITIONAL INVESTIGATIONAL INPUTS WERE AVAILABLE. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
PATIENT WAS REVISED TO ADDRESS PAIN, TIBIAL SUBSIDENCE AND LOOSENING AT BOTH INTERFACES. THE CEMENT MANUFACTURER IS UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 722158 | PFC SIGMA MOD TIB TRAY CEM SZ2 | KNEE TIBIAL TRAY | JWH | DEPUY IRELAND 9616671 | 1058701 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Required Intervention |