X3 TRIATHLON CS INS SIZE 4 9MM
Report
- Report Number
- 0002249697-2014-04107
- Event Type
- Malfunction
- Date Received
- November 3, 2014
- Date of Event
- October 8, 2014
- Report Date
- October 8, 2014
- Manufacturer
- STRYKER ORTHOPAEDICS-MAHWAH
- Product Code
- JWH
- PMA / PMN Number
- K063423
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PHYSICIAN
Narratives
THE PATIENT IS (B)(6) CENTIMETERS IN HEIGHT. AN EVENT REGARDING A SEATING ISSUE INVOLVING A TRIATHLON INSERT WAS REPORTED. THE EVENT WAS CONFIRMED. METHOD & RESULTS: DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION OF THE RETURNED COMPONENT INDICATES THAT THE INSERT MAY HAVE BEEN OBSTRUCTED DURING THE ATTEMPTED SEATING AND THAT THE INSERT WAS AT LEAST PARTIALLY ASSEMBLED WITH THE BASEPLATE. DEVICE HISTORY REVIEW: DHR REVIEW FOR THE REPORTED LOT WAS SATISFACTORY. COMPLAINT HISTORY REVIEW: CHR REVIEW FOR THE REPORTED LOT CONFIRMED THAT THERE ARE NO OTHER SIMILAR EVENTS REPORTED. CONCLUSIONS: BASED ON THE VISUAL INSPECTION OF THE RETURNED COMPONENT, THE INSERT MAY HAVE BEEN OBSTRUCTED DURING THE ATTEMPTED SEATING AND THE INSERT WAS AT LEAST PARTIALLY ASSEMBLED WITH THE BASEPLATE. THE DAMAGE OBSERVED ON THE PART WAS CAUSED BY THE USER AND WOULD HAVE RENDERED THE COMPONENT UNUSABLE.
A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
LEFT TOTAL KNEE REPLACEMENT (TRIATHLON). SURGEON IMPLANTED A CEMENTED SIZE 4 TIBIAL BASEPLATE, A SIZE 4 PA FEMUR AND A CEMENTED 29MM ASYMMETRIC PATELLA. HE CHOSE TO USE A SIZE 4 9MM CS INSERT AND WHEN HE IMPACTED IT, IT WOULDN'T SEAT PROPERLY ON THE BASEPLATE. THE CS INSERT "BOUNCED" AND WAS SPRINGY ON THE MEDIAL SIDE AND WAS FLUSH ON THE LATERAL SIDE WHEN FULLY IMPACTED. THERE WAS NO TISSUE IMPINGEMENT SEEN. THE SURGEON WAS UNABLE TO SEAT THE INSERT ANY FURTHER, EVEN AFTER MORE IMPACTION. HE DECIDED TO REMOVE THIS CS INSERT AND REPLACE IT WITH A SIZE 4 9MM CR INSERT.
LEFT TOTAL KNEE REPLACEMENT (TRIATHLON). SURGEON IMPLANTED A CEMENTED SIZE 4 TIBIAL BASEPLATE, A SIZE 4 PA FEMUR AND A CEMENTED 29MM ASYMMETRIC PATELLA. HE CHOSE TO USE A SIZE 4 9MM CS INSERT AND WHEN HE IMPACTED IT WOULDN'T SEAT PROPERLY ON THE BASEPLATE. THE CS INSERT "BOUNCED" AND WAS SPRINGY ON THE MEDIAL SIDE AND WAS FLUSH ON THE LATERAL SIDE WHEN FULLY IMPACTED. THERE WAS NO TISSUE IMPINGEMENT SEEN. THE SURGEON WAS UNABLE TO SEAT THE INSERT ANY FURTHER, EVEN AFTER MORE IMPACTION. HE DECIDED TO REMOVE THIS CS INSERT AND REPLACE IT WITH A SIZE 4 9MM CR INSERT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 701197 | X3 TRIATHLON CS INS SIZE 4 9MM | IMPLANT | JWH | STRYKER ORTHOPAEDICS-MAHWAH | LDT842 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | Other |