X3 TRIATHLON INSERT PS#7 16MM
Report
- Report Number
- 0002249697-2014-02295
- Event Type
- Malfunction
- Date Received
- June 16, 2014
- Date of Event
- May 25, 2014
- Report Date
- May 25, 2014
- Manufacturer
- STRYKER ORTHOPAEDICS-MAHWAH
- Product Code
- JWH
- PMA / PMN Number
- K051146
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NZ
- Reporter Occupation
- OTHER
Narratives
A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
AN EVENT REGARDING DAMAGE INVOLVING A X3 TRIATHLON INSERT PS#7 16MM WAS REPORTED. THE EVENT WAS CONFIRMED. THE LOCKING WIRE OF THE INSERT IS DISENGAGED FROM THE POLY AND THERE ARE MINOR SCRATCHES ON THE BODY OF THE DEVICE. DIMENSIONAL AND FUNCTIONAL EVALUATIONS WERE NOT PERFORMED AS THE DEVICE IS DAMAGED. MEDICAL EVALUATION NOT PERFORMED AS NO MEDICAL RECORDS WERE PROVIDED. DEVICE HISTORY REVIEW INDICATED THAT THE SPECIFIED LOT WAS ACCEPTED INTO STOCK WITH NO REPORTED DISCREPANCIES. COMPLAINT HISTORY REVIEW INDICATED THAT THERE HAVE NOT BEEN ANY OTHER EVENTS FOR THE SPECIFIED LOT. THE REPORTED EVENT INVOLVES THE LOCKING WIRE ON AN INSERT BEING NOT PROPERLY ENGAGED. THE INSERT WAS DISCARDED AND A NEW INSERT WAS IMPLANTED. IT IS LIKELY THAT THE EVENT IS A USER RELATED ISSUE WITH INSERTING THE DEVICE PROPERLY. NO FURTHER INVESTIGATION FOR THIS EVENT IS POSSIBLE AT THIS TIME.
CUSTOMER NOTICED THE LOCKING WIRE ON A PS INSERT WAS NOT PROPERLY ENGAGED. THE SURGEON DISCARDED THIS ITEM AND INSERTED ANOTHER ONE.
CUSTOMER NOTICED THE LOCKING WIRE ON A PS INSERT WAS NOT PROPERLY ENGAGED. THE SURGEON DISCARDED THIS ITEM AND INSERTED ANOTHER ONE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 352049 | X3 TRIATHLON INSERT PS#7 16MM | IMPLANT | JWH | STRYKER ORTHOPAEDICS-MAHWAH | LCT316 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |