TRIMA ACCEL
Report
- Report Number
- 1722028-2014-00227
- Event Type
- Malfunction
- Date Received
- June 13, 2014
- Date of Event
- May 17, 2014
- Report Date
- May 19, 2014
- Manufacturer
- TERUMO BCT
- Product Code
- GKT
- PMA / PMN Number
- BK120049
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
INVESTIGATION: THE DEVICE HISTORY RECORDS (DHR) WERE REVIEWED FOR THIS EVENT. THERE WERE NO EVENTS NOTED IN THE DHR THAT WOULD HAVE CONTRIBUTED TO THE ELEVATED WBC COUNT EXPERIENCED BY THE CUSTOMER. THE RUN DATA FILE (RDF) WAS ANALYZED FOR THIS EVENT. ROOT CAUSE: A DEFINITIVE ROOT CAUSE FOR THE OBSERVED LEUKOREDUCTION FAILURE REMAINS UNDETERMINED AT THIS TIME. THE SIGNALS IN THE RDF INDICATE THAT THE HIGHER-THAN-EXPECTED WBC CONTENT IN THE PLATELET PRODUCT WAS LIKELY A RESULT OF AN ESCAPE OF WBCS FROM THE LRS CHAMBER DURING A PORTION OF THE PROCEDURE. THERE ARE NO EVENTS (ADJUSTMENTS,CHANGES IN PUMP SPEED, SUBSTATE CHANGES, ETC.) IN THE PROCEDURE THAT CORRESPONDS WITH THE ONSET OF THE OVERLOADING OF THE LRS CHAMBER. BASED ON THE AVAILABLE INFORMATION, IT CANNOT BE RULED OUT THAT THE HIGHER-THAN-EXPECTED WBC CONTENT IN THE PLATELET PRODUCT COULD BE DONOR-RELATED. IT ALSO CANNOT BE RULED OUT THAT A SAMPLING, CALCULATION, OR OTHER PROCESS ERROR COULD HAVE CONTRIBUTED TO THE HIGHER-THAN-EXPECTED WBC CONTENT IN THE PLATELET PRODUCT.
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE PLATELET PRODUCT. THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RWBC TESTING,THEREFORE, NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. DONOR UNIT# (B)(6) THE DISPOSABLE KIT IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER. THIS REPORT IS BEING FILED DUE TO A DEVICE MALFUNCTION, THAT HAS THE POTENTIAL FOR INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 350388 | TRIMA ACCEL | TRIMA ACCEL PLATELET, PLASMA SET | GKT | TERUMO BCT | 03W2213 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |