FDA Adverse Event Malfunction Summary report: N

TRIATHLON PRIM TIB BASEPLATE - CEMENTED

MDR report key: 3825275 · Received May 21, 2014

Report

Report Number
0002249697-2014-01907
Event Type
Malfunction
Date Received
May 21, 2014
Date of Event
April 24, 2014
Report Date
April 28, 2014
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
JWH
PMA / PMN Number
K031729
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

AN EVENT REGARDING PACKAGING DAMAGE INVOLVING A TRIATHLON BASEPLATE WAS REPORTED. THE EVENT WAS CONFIRMED. THE RETURNED OUTER BLISTER HAS THE LIP OF ONE SIDE FULLY BROKEN OFF. NO OTHER PACKAGING WAS RETURNED. DEVICE HISTORY REVIEW FOR THE REPORTED LOT DETERMINED THAT THE DEVICE WAS MANUFACTURED AND PACKED TO SPECIFICATION. COMPLAINT HISTORY REVIEW DETERMINED THAT THERE WERE NO SIMILAR EVENTS REPORTED FOR THE LOT. THE RETURNED OUTER BLISTER HAS THE LIP OF ONE SIDE FULLY BROKEN OFF. NO OTHER PACKAGING WAS RETURNED FOR ANALYSIS. THE COMPLAINT DESCRIPTION STATES "THAT PART OF THE PLASTIC PACKAGE LEFT IN THE LABEL WHEN PACKAGE WAS OPENED". ONE SIDE OF THE BLISTER HAD THE FLANGE/LIP BROKEN OFF AND REMAINED ATTACHED TO THE TYVEK. THE EVENT DESCRIPTION REFERS TO THE TYVEK LID AS THE LABEL. INSPECTION OF THE ENTIRE PACKAGING OF THE PRODUCT IS REQUIRED TO DETERMINE THE ROOT CAUSE OF THIS EVENT. IT DEVICES AND/OR ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS INVESTIGATION WILL BE REOPENED. NO FURTHER INVESTIGATION FOR THIS EVENT IS POSSIBLE AT THIS TIME.

Additional Manufacturer Narrative · 1

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

NURSE INFORMED SALES REP THAT PART OF THE PLASTIC PACKAGE LEFT IN THE LABEL WHEN PACKAGE WAS OPENED. WHEN THE PACKAGE WAS OPENED, ONE OF THE EDGE OF THE PLASTIC BOX WAS STUCK TO THE LABEL. THE INNER PACKAGING WAS INTACT. CUSTOMER USED THE PRODUCT. OPERATION TIME DID NOT GET LONGER.

Description of Event or Problem · 1

NURSE INFORMED SALES REP THAT PART OF THE PLASTIC PACKAGE LEFT IN THE LABEL WHEN PACKAGE WAS OPENED. WHEN THE PACKAGE WAS OPENED, ONE OF THE EDGE OF THE PLASTIC BOX WAS STUCK TO THE LABEL. THE INNER PACKAGING WAS INTACT. CUSTOMER USED THE PRODUCT. OPERATION TIME DID NOT GET LONGER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303057 TRIATHLON PRIM TIB BASEPLATE - CEMENTED IMPLANT JWH STRYKER ORTHOPAEDICS-MAHWAH EH9HA

Patients

Seq Age Sex Outcome Treatment
1 Other