PFC*CVD TB INS,SZ3, 12.5MM

MDR report key: 3800441 · Received May 8, 2014

Report

Report Number
1818910-2014-18361
Event Type
Injury
Date Received
May 8, 2014
Date of Event
April 17, 2014
Report Date
June 5, 2014
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
JWH
PMA / PMN Number
PK884796
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A. PATIENT MEDICAL RECORDS WERE PROVIDED. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A LOT SPECIFIC COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE FOR THE PRODUCT LOT CODE REQUIRED WAS NOT PROVIDED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. DUE TO THE FACT THAT THE PATIENT WAS MORBIDLY OBESE COUPLED WITH THE LENGTH OF TIME BETWEEN DATE OF INSERTION AND THE DATE OF REVISION, IT WOULD NOT BE UNEXPECTED TO OBSERVE POLY WEAR. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS INSTABILITY RESULTING FROM POLY WEAR. PAIN WAS ALSO REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
277554 PFC*CVD TB INS,SZ3, 12.5MM KNEE TIBIAL BEARING/INSERT JWH DEPUY ORTHOPAEDICS, INC. UNK

Patients

Seq Age Sex Outcome Treatment
1 89 YR Required Intervention