FDA Adverse Event Malfunction Summary report: N

PATELLA CALIPER

MDR report key: 3781399 · Received April 30, 2014

Report

Report Number
0002249697-2014-01574
Event Type
Malfunction
Date Received
April 30, 2014
Date of Event
April 7, 2014
Report Date
April 7, 2014
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
JWH
PMA / PMN Number
K123486
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Additional Manufacturer Narrative · 1

AN EVENT REGARDING DEGRADED HANDLES INVOLVING A TRIATHLON PATELLAR CALIPER WAS REPORTED. THE EVENT WAS CONFIRMED. METHOD & RESULTS: -DEVICE EVALUATION AND RESULTS: VISUAL ANALYSIS CONFIRMED THE REPORTED EVENT. -MEDICAL RECORDS RECEIVED AND EVALUATION: NOT PERFORMED BECAUSE IT IS BELIEVED PATIENT FACTORS DID NOT CONTRIBUTE TO THE REPORTED EVENT. - DEVICE HISTORY REVIEW INDICATED ALL DEVICES ACCEPTED INTO STOCK MET SPECIFICATION. - COMPLAINT HISTORY REVIEW INDICATED THERE HAVE BEEN TWO OTHER EVENTS ASSOCIATE WITH THE REPORTED LOT. CONCLUSION: CAPA WAS INITIATED BECAUSE A SERIES OF COMPLAINTS WERE RECEIVED FOR TRIATHLON INSTRUMENTATION WHERE GREEN SANTOPRENE OVER-MOLD HANDLES HAVE BEEN REPORTED TO BE DEGRADING, BECOMING STICKY AND UNSTABLE. THIS EVENT WAS DETERMINED TO BE UNDER THE SCOPE OF CAPA. CHEMICAL AND CYTOTOXICITY TESTS SHOWED THE DEGRADED SANTOPRENE IS NON-TOXIC. THE ABILITY TO CLEAN AND STERILIZE THE DEGRADED INSTRUMENTS HAS NOT BEEN COMPROMISED.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE CHECKING INSTRUMENTS I NOTICED THE RUBBER HANDLES ON THE PATELLA CLAMP, AND THE PATELLA CALIPER WERE STICKY AND THE RUBBER APPEARS TO BE UNSTABLE AND EASILY SCRAPED WITH A FINGER NAIL.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE CHECKING INSTRUMENTS I NOTICED THE RUBBER HANDLES ON THE PATELLA CLAMP, AND THE PATELLA CALIPER WERE STICKY AND THE RUBBER APPEARS TO BE UNSTABLE AND EASILY SCRAPED WITH A FINGER NAIL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
260671 PATELLA CALIPER INSTRUMENT JWH STRYKER ORTHOPAEDICS-MAHWAH SB2V89

Patients

Seq Age Sex Outcome Treatment
1 Other