FDA Adverse Event Injury Summary report: N

NEXGEN LPS-FLEX OPTION FEMORAL COMPONENT

MDR report key: 3522766 · Received December 5, 2013

Report

Report Number
1822565-2013-01744
Event Type
Injury
Date Received
December 5, 2013
Date of Event
September 23, 2009
Report Date
March 17, 2020
Manufacturer
ZIMMER INC
Product Code
JWH
PMA / PMN Number
K991581
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INFORMATION WAS RECEIVED FROM A CONSUMER WHO IS NOT REQUIRED TO COMPLETE FORM 3500A. THIS REPORT WILL BE AMENDED WHEN OUR INVESTIGATION IS COMPLETE.

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY UPDATED AND ADDITIONAL INFORMATION. MEDICAL PRODUCTS: LPS-FLEX OPTION FEMORAL G-L; P/N: 00596401701, L/N: 54769200; NEXGEN LPS-MOBILE ARTICULAR SURFACE SIZE G 14MM; P/N: 99594607014, L/N: 70801400; OSTEOBOND BONE CEMENT; P/N: 00110100800, L/N: 5582900; ALL POLY PAT COMP 32DIA; P/N: 00597206532, L/N: 60003654; NEXGEN FLUTED STEM MOBILE TIBIAL COMPONENT SIZE 6; P/N: 99594606001, L/N: 69020500. COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS CONFIRMED. VISUAL EVALUATION OF THE RETURNED FEMUR AND ARTICULAR SHOWS DAMAGES RELATED TO USE LIKE NICKS, GOUGES, WEARS. THE FEMUR DIMENSIONS WERE WITHIN THE SPECIFICATION OF THE PRINT, WHERE MEASURED. DHR WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. PER PACKAGE LOOSENING, PAIN, AND INSTABILITY ARE KNOWN ADVERSE EFFECTS OF THIS SYSTEM. WITH UPDATED INFORMATION, NO CHANGE IN THE ROOT CAUSE. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 1

IT IS REPORTED THAT THE PATIENT WAS REVISED DUE TO PAIN.

Description of Event or Problem · 1

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
631692 NEXGEN LPS-FLEX OPTION FEMORAL COMPONENT JWH ZIMMER INC 54769200

Patients

Seq Age Sex Outcome Treatment
1 54 YR Hospitalization| R