FDA Adverse Event Injury Summary report: N

GMK FIXED CEMENTED TIBIAL TRAY

MDR report key: 3519927 · Received November 22, 2013

Report

Report Number
3005180920-2013-00146
Event Type
Injury
Date Received
November 22, 2013
Manufacturer
MEDACTA INTERNATIONAL, SA
Product Code
JWH
PMA / PMN Number
K090988
Report Source
Manufacturer report

Narratives

Additional Manufacturer Narrative · 1

DOCUMENT REVIEW: GMK FIXED CEMENTED TIBIAL TRAY SIZE 3 RIGHT. CODE (B)(4) / LOT 100866 (34 ITEMS PRODUCED): ALL PARAMETERS WERE FOUND TO BE IN ACCORDANCE WITH THE SPECIFICATIONS VALID AT THE TIME OF MANUFACTURING. ALL THE ITEMS BELONGING TO THIS LOT HAVE BEEN ALREADY SOLD AND NO SIMILAR INCIDENTS HAVE BEEN REPORTED. ON THE BASIS OF THE DATA COLLECTED, WE HAVE NO EVIDENCES THAT THE PROBLEM IS DEVICE RELATED, BUT IT IS LIKELY RELATED TO A NOT PROPER CEMENTATION TECHNIQUE DURING THE PRIMARY SURGERY.

Description of Event or Problem · 1

REFERENCE IMPORTER REPORT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
609372 GMK FIXED CEMENTED TIBIAL TRAY CEMENTED TIBIAL BASE PLATE JWH MEDACTA INTERNATIONAL, SA

Patients

Seq Age Sex Outcome Treatment
1