FDA Adverse Event Injury Summary report: N

NEXGEN LPS-FLEX PROLONG ARTICULAR SURFACE

MDR report key: 3434352 · Received October 24, 2013

Report

Report Number
1822565-2013-01545
Event Type
Injury
Date Received
October 24, 2013
Date of Event
September 3, 2013
Report Date
September 26, 2013
Manufacturer
ZIMMER, INC.
Product Code
JWH
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SURGICAL NOTES WERE PROVIDED FOR REVIEW. THE PRIMARY OPERATIVE NOTES STATE THAT WITH TRIAL COMPONENTS, EXCELLENT MOBILITY AND TRACKING WAS ACHIEVED. THE REVISION NOTES CONFIRMED THAT THE POST HAD FRACTURED REVISION NOTES CONFIRMED THAT THE POST HAD FRACTURED AND WAS FOUND WITHIN THE INTER-CONDYLAR NOTCH. THE FRACTURED PIECE AND THE REMAINING ARTICULAR SURFACE WERE EXCHANGED. X-RAYS WERE NOT PROVIDED; IT IS UNK WHETHER THE COMPONENTS WERE IMPLANTED WITH THE CORRECT FIT AND ORIENTATION AS PER THE SURGICAL TECHNIQUE. REHABILITATION PROTOCOL AND ADHERENCE THERETO IS UNK. REVIEW OF COMPLAINT HISTORY FOUND NO OTHER REPORTS OF THIS NATURE. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED WITH THE INFO PROVIDED. EVAL CODES, REVIEW OF THE DEVICE HISTORY RECORDS DID NOT FIND ANY DEVIATIONS OR ANOMALIES. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADD'L SUBSTANTIVE INFO BE RECEIVED, THE COMPLAINT WILL BE REOPENED. ZIMMER, INC. CONSIDERS THE INVESTIGATION CLOSED.

Description of Event or Problem · 1

IT IS REPORTED THAT THE PT WAS REVISED DUE TO MILD SWELLING, CLICKING OR POPPING, AND MILD HYPER EXTENSION. THE ARTICULAR SURFACE POST WAS FRACTURED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
546204 NEXGEN LPS-FLEX PROLONG ARTICULAR SURFACE JWH ZIMMER, INC. 60637917

Patients

Seq Age Sex Outcome Treatment
1 59 YR Required Intervention