FDA Adverse Event Malfunction Summary report: N

TRIMA ACCEL

MDR report key: 3261374 · Received August 2, 2013

Report

Report Number
1722028-2013-01328
Event Type
Malfunction
Date Received
August 2, 2013
Date of Event
July 8, 2013
Report Date
July 8, 2013
Manufacturer
TERUMO BCT
Product Code
GKT
PMA / PMN Number
BK120017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: THE DISPOSABLE SET WAS UNAVAILABLE FOR RETURN AND INVESTIGATION. THE RUN DATA FILE WAS ANALYZED FOR THIS EVENT. THE ANALYSIS OF THE RUN DATA FILE DID NOT FIND A CONCLUSIVE CAUSE FOR THE HIGHER-THAN-EXPECTED WBC CONTENT IN THE PLATELET PRODUCT REPORTED FOR THIS COLLECTION. NO UNUSUAL PROCESS VARIABLE WAS IDENTIFIED AND THE SIGNALS IN THE RUN DATA FILE INDICATE THAT THE TRIMA ACCEL SYSTEM OPERATED AS INTENDED. BASED ON THE AVAILABLE INFORMATION, IT CANNOT BE RULED OUT THAT THE HIGHER THAN EXPECTED WBC CONTENT IN THE PLATELET PRODUCT COULD BE DONOR-RELATED. IT ALSO CANNOT BE RULED OUT THAT A SAMPLING, CALCULATION, OR OTHER PROCESS ERROR COULD HAVE CONTRIBUTED TO THE HIGHER THAN EXPECTED WBC CONTENT IN THE PLATELET PRODUCT. INVESTIGATION EVALUATION AND CORRECTIVE ACTIONS ARE IN-PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED.

Additional Manufacturer Narrative · 1

INVESTIGATION: THE DEVICE HISTORY RECORDS (DHR) WERE REVIEWED FOR THIS EVENT. THERE WERE NO EVENTS NOTED IN THE DHR THAT WOULD HAVE CONTRIBUTED TO THE ELEVATED WBC COUNT EXPERIENCED BY THE CUSTOMER. ROOT CAUSE: A DEFINITIVE ROOT CAUSE FOR THE OBSERVED LEUKO REDUCTION FAILURE REMAINS UNDETERMINED AT THIS TIME. POSSIBLE ROOT CAUSES WERE PROVIDED IN THE INITIAL REPORT FOR THIS EVENT.

Description of Event or Problem · 1

THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE PLATELET PRODUCT. THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. DONOR UNIT #: (B)(6). THE DISPOSABLE SET IS UNAVAILABLE FOR RETURN BECAUSE THE CUSTOMER DISCARDED IT. THIS REPORT IS BEING FILED DUE TO A DEVICE MALFUNCTION THAT HAS THE POTENTIAL FOR INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
362801 TRIMA ACCEL TRIMA ACCEL ENHANCED PLATELET, PLASMA, RBC SET GKT TERUMO BCT 04V1118

Patients

Seq Age Sex Outcome Treatment
1 Other