FDA Adverse Event Malfunction Summary report: N

TRIMA ACCEL

MDR report key: 3111517 · Received May 14, 2013

Report

Report Number
1722028-2013-01192
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
April 15, 2013
Report Date
April 16, 2013
Manufacturer
TERUMO BCT
Product Code
GKT
PMA / PMN Number
BK120017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: THE RUN DATA FILE WAS ANALYZED FOR THIS EVENT. ROOT CAUSE: A DEFINITIVE ROOT CAUSE FOR THE OBSERVED LEUKOREDUCTION FAILURE REMAINS UNDETERMINED AT THIS TIME. SIGNALS IN THE RUN DATA FILE DO NOT INDICATE A CONCLUSIVE CAUSE FOR THE GREATER THAN EXPECTED RWBC CONTENT IN THE PLATELET PRODUCT REPORTED FOR THIS COLLECTION. NO UNUSUAL PROCESS VARIABLE WAS IDENTIFIED IN THE RUN DATA FILE AND THE TRIMA ACCEL SYSTEM OPERATED AS INTENDED. BASED ON THE AVAILABLE INFORMATION IT IS POSSIBLE THAT THIS LEUKOREDUCTION FAILURE COULD DONOR-RELATED. IT ALSO CANNOT BE RULED OUT THAT A SAMPLING, CALCULATION, OR OTHER PROCESS ERROR MAY HAVE CONTRIBUTED TO THE HIGHER THAN EXPECTED RWBC CONTENT IN THE PLATELET PRODUCT.

Description of Event or Problem · 1

THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE PLATELET PRODUCT. THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. DONOR UNIT #: (B)(6) THE DISPOSABLE SET IS UNAVAILABLE FOR RETURN BECAUSE THE CUSTOMER DISCARDED IT. THIS REPORT IS BEING FILED DUE TO A DEVICE MALFUNCTION THAT HAS THE POTENTIAL FOR INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
213050 TRIMA ACCEL TRIMA ACCEL PLATELET, PLASMA, RBC SET GKT TERUMO BCT 02V2226

Patients

Seq Age Sex Outcome Treatment
1 Other