TRIMA ACCEL
Report
- Report Number
- 1722028-2013-01140
- Date Received
- April 23, 2013
- Date of Event
- February 27, 2013
- Report Date
- March 27, 2013
- Manufacturer
- TERUMO CORPORATION/TERUMO BCT
- Product Code
- GKT
- PMA / PMN Number
- BK120017
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THIS REPORT IS BEING FILED TO PROVIDE ADDITIONAL INFORMATION. INVESTIGATION: A REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR THIS UNIT SHOWED NO IRREGULARITIES DURING MANUFACTURING THAT WERE RELEVANT TO THIS ISSUE. ROOT CAUSE: THIS DISPOSABLE SET WAS UNAVAILABLE FOR SPECIFIC ROOT CAUSE ANALYSIS. BACTERIAL CONTAMINATION CAN OCCUR AT ANY POINT ALONG THE PROCESS OF COLLECTING PRODUCTS. POSSIBLE CAUSES INCLUDE BUT ARE NOT LIMITED TO:-IMPROPER CLEANSING AND/OR DONOR PREP-NOT DIVERTING THE FIRST 10ML OF DONOR BLOOD-IMPROPER STORAGE-UNDETECTED MISASSEMBLY IN THE SET
INVESTIGATION EVALUATION AND CORRECTIVE ACTIONS ARE IN-PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED
THIS REPORT IS BEING FILED TO PROVIDE ADDITIONAL INFORMATION.
NO MEDICAL INTERVENTION WAS NECESSARY FOR THIS EVENT.
THE CUSTOMER REPORTED STERILITY TESTING IN THE QUALITY CONTROL LABORATORY REVEALED POSITIVE CONTAMINATION. THE CUSTOMER WAS NOT WILLING TO PROVIDE THE PATIENT IDENTIFIER. THE DISPOSABLE SET IS UNAVAILABLE FOR RETURN FOR EVALUATION. THIS REPORT IS BEING FILED DUE TO INSUFFICIENT INFORMATION PROVIDED AT THIS TIME TO DETERMINE IF MALFUNCTION WITH THE POTENTIAL FOR INJURY OCCURRED.
THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THIS INCIDENT,THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN. PATIENT INFORMATION: NONE THE DONOR INFORMATION WAS PROVIDED ORIGINALLY IN ORDER TO ALIGN WITH THE COMPLAINT INVESTIGATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 174651 | TRIMA ACCEL | TRIMA ACCEL PLT, PLS, RBC AUTO PAS SET | GKT | TERUMO CORPORATION/TERUMO BCT | 11U1216 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00025 YR |