FDA Adverse Event Malfunction Summary report: N

COBE SPECTRA BLOOD COLLECTION

MDR report key: 3031351 · Received April 2, 2013

Report

Report Number
1722028-2013-01068
Event Type
Malfunction
Date Received
April 2, 2013
Date of Event
March 7, 2013
Report Date
March 7, 2013
Manufacturer
TERUMO BCT
Product Code
GKT
PMA / PMN Number
K893962
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: A PRODUCT DISPOSITION FORM REVIEW WAS COMPLETED FOR THIS MACHINE SERIAL NUMBER AND NO MANUFACTURING-RELATED ISSUES WERE FOUND. INVESTIGATION EVALUATION AND CORRECTIVE ACTIONS ARE IN-PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING FILED TO PROVIDE ADDITIONAL INFORMATION. INVESTIGATION: NO FUNCTIONAL CHECKOUT OF THE DEVICE WAS PERFORMED AS NO MALFUNCTION OF THE DEVICE WAS ALLEGED. PER CONVERSATIONS WITH THE CUSTOMER, THIS INCIDENT IS ATTRIBUTED TO AN OPERATOR DATA ENTRY ERROR. ROOT CAUSE: OPERATOR ERROR AN INTERNAL ISSUE TRACKER HAS BEEN CREATED TO INVESTIGATE POSSIBLE MITIGATIONS FOR THE POSSIBILITY OF AN OPERATOR ENTERING A LOWER THAN PRESCRIBED HEMATOCRIT.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT APPROXIMATELY HALFWAY THROUGH A RED BLOOD CELL EXCHANGE PROCEDURE, THE OPERATOR REALIZED HE HAD ENTERED THE INCORRECT REPLACEMENT FLUID HEMATOCRIT(HCT). THE OPERATOR HAD ACCEPTED THE DEFAULT HCT OF 80%, BUT THEY WERE USING UNITS WITH A 55% HCT. THE TERUMO BCT SUPPORT SPECIALIST SUGGESTED TO THE CUSTOMER TO END THE PROCEDURE, GET THE PATIENT'S HCT, AND CONTACT THE PHYSICIAN. THE PATIENT WAS STABLE FOLLOWING THE PROCEDURE. PATIENT IDENTIFIER AND AGE ARE NOT AVAILABLE AT THIS TIME. THIS REPORT IS BEING FILED DUE TO DEVICE MALFUNCTION, IN THE FORM OF OPERATOR ERROR, THAT HAS THE POTENTIAL FOR INJURY.

Description of Event or Problem · 1

THE CUSTOMER COULD NOT PROVIDE THE PATIENT'S IDENTIFIER OR AGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
134138 COBE SPECTRA BLOOD COLLECTION COBE SPECTRA APHERESIS SYSTEM GKT TERUMO BCT 000000000950000000

Patients

Seq Age Sex Outcome Treatment
1 Other