FDA Adverse Event Malfunction Summary report: N

TRIATHLON #4 PS INSERT 13MM

MDR report key: 3030319 · Received April 2, 2013

Report

Report Number
0002249697-2013-01159
Event Type
Malfunction
Date Received
April 2, 2013
Date of Event
March 11, 2013
Report Date
March 13, 2013
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
JWH
PMA / PMN Number
K031729
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

WHEN COMPLETED, THE EVALUATION SUMMARY WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

BASED ON THE VISUAL INSPECTION IT APPEARS THAT THE IMPLEMENT USED TO PUSH BACK ON THE ANTERIOR FACE OF THE INSERT COMPONENT DURING ATTEMPTED SEATING MAY HAVE COME INTO CONTACT WITH THE LOCKING WIRE. WITHOUT THE RETURN OF THE LOCKING WIRE FOR EVALUATION, IT IS NOT POSSIBLE TO DEFINITIVELY DETERMINE THE ROOT CAUSE. THE EVENT AS REPORTED WAS CONFIRMED. DHR REVIEW FOR THE REPORTED LOT WAS SATISFACTORY. CHR REVIEW FOR THE REPORTED LOT CONFIRMED THAT THERE WERE NO SIMILAR EVENTS REPORTED FOR THE LOT.

Description of Event or Problem · 1

IT WAS REPORTED ON BEHALF OF THE CUSTOMER THAT DURING SURGERY WHEN THE SURGEON WAS INTRODUCING THE INSERT INTO THE TIBIAL BASEPLATE, THE LOCKING WIRE BROKE. ACCORDING TO THE SURGEON NO STRESS WAS USED WHEN INTRODUCING THE INSERT ONLY GENTLE PRESSURE WITH A FINGER AND INSERT IMPACTOR. ANOTHER INSERT WAS USED. NO DELAYS OR ADVERSE CONSEQUENCES WERE REPORTED.

Description of Event or Problem · 1

IT WAS REPORTED ON BEHALF OF THE CUSTOMER THAT DURING SURGERY WHEN THE SURGEON WAS INTRODUCING THE INSERT INTO THE TIBIAL BASEPLATE, THE LOCKING WIRE BROKE. ACCORDING TO THE SURGEON NO STRESS WAS USED WHEN INTRODUCING THE INSERT ONLY GENTLE PRESSURE WITH A FINGER AND INSERT IMPACTOR. ANOTHER INSERT WAS USED. NO DELAYS OR ADVERSE CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
133627 TRIATHLON #4 PS INSERT 13MM IMPLANT JWH STRYKER ORTHOPAEDICS-MAHWAH LCG319

Patients

Seq Age Sex Outcome Treatment
1 Other