FDA Adverse Event Summary report: N

3DKNEE

MDR report key: 3012078 · Received March 20, 2013

Report

Report Number
1644408-2013-00161
Date Received
March 20, 2013
Date of Event
March 11, 2013
Report Date
March 11, 2013
Manufacturer
ENCORE MEDICAL, L.P.
Product Code
JWH
PMA / PMN Number
K020114
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REASON FOR THIS REVISION SURGERY WAS REPORTED AS WEAR AFTER 7.9 YEARS OF PATIENT USE. THERE IS NO INFORMATION IN THIS COMPLAINT ABOUT ANY PATIENT INJURIES, ACTIVITIES, ACCIDENTS, OR MEDICAL CONTRAINDICATIONS THAT MAY HAVE CONTRIBUTED TO THE REVISION SURGERY. THERE WAS NO DELAY IN SURGERY AND ANOTHER SUITABLE DEVICE WAS AVAILABLE FOR USE. THE REVISION SURGERY WAS COMPLETED AS INTENDED. THE DEVICE WAS NOT RETURNED TO DJO SURGICAL FOR EXAMINATION. A REVIEW OF THE DEVICE HISTORY RECORDS SHOWED NO NON-CONFORMING MATERIAL REPORTS ASSOCIATED WITH THIS PRODUCT. A REVIEW OF THE PRODUCT COMPLAINT REPORT HISTORY SHOWS THIS IS THE THIRD COMPLAINT FOR THIS PART NUMBER; ONE DUE TO WEAR AND TWO REVISIONS. THE ROOT CAUSE FOR THE REVISION WAS DUE TO WEAR. THERE IS NO INFORMATION REPORTED THAT SHOWED A MATERIAL, DESIGN, OR MANUFACTURING PROBLEM WITH THE PRODUCT. THERE ARE NO INDICATIONS OF A PRODUCT OR PROCESS ISSUE AFFECTING IMPLANT SAFETY OR EFFECTIVENESS.

Description of Event or Problem · 1

REVISION SURGERY - THE POLY INSERT HAD SIGNIFICANT WEAR ON THE LATERAL SIDE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
116623 3DKNEE INSERT SIZE 10, 9MM, LEFT CM JWH ENCORE MEDICAL, L.P. 416042

Patients

Seq Age Sex Outcome Treatment
1 58 YR Required Intervention