FDA Adverse Event Malfunction Summary report: N

NEXGEN LPS-FLEX ARTICULAR SURFACE

MDR report key: 3002374 · Received March 7, 2013

Report

Report Number
1822565-2013-00450
Event Type
Malfunction
Date Received
March 7, 2013
Date of Event
February 5, 2013
Report Date
February 5, 2013
Manufacturer
ZIMMER, INC.
Product Code
JWH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL SUMMARY: THIS TYPICALLY HAPPENS WHEN THE ARTICULAR SURFACE IS NOT OPTIMALLY PLACED AND ORIENTED BEFORE ATTEMPTED INSERTION USING THE ARTICULAR SURFACE INSERTER INSTRUMENT. SUBOPTIMAL ORIENTATION PRIOR TO ATTEMPTED INSERTION APPEARS TO BE THE MOST LIKELY CAUSE OF FAILURE. AS RETURNED, THE DOVETAIL IS DEFORMED. MEASURED DIMENSIONS ARE WITHIN PRINT SPECIFICATIONS. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL SUBSTANTIVE INFO BE RECEIVED, THE COMPLAINT WILL BE REOPENED. ZIMMER, INC. CONSIDERS THE INVESTIGATION CLOSED.

Description of Event or Problem · 1

IT IS REPORTED THAT THE SURGEON WAS UNABLE TO PROPERLY SEAT THE ARTICULAR SURFACE. ANOTHER SURFACE OF THE SAME SIZE AND LOT WAS USED TO COMPLETE THE SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
97067 NEXGEN LPS-FLEX ARTICULAR SURFACE JWH ZIMMER, INC. 62278653

Patients

Seq Age Sex Outcome Treatment
1