Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: NEITHER SURGICAL NOTES NOR X-RAYS WERE PROVIDED, THEREFORE, FIT AND ORIENTATION COULD NOT BE EVALUATED. PT FACTORS THAT MAY AFFECT THE PERFORMANCE OF THE COMPONENTS SUCH AS BONE QUALITY, ACTIVITY LEVEL OR TYPE OF ACTIVITY (LOW IMPACT VS. HIGH IMPACT) ARE UNK. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED WITH THE INFORMATION PROVIDED. EVALUATION: REVIEW OF THE DEVICE HISTORY RECORDS DID NOT FIND ANY DEVIATIONS OR ANOMALIES. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL SUBSTANTIVE INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. ZIMMER, INC. CONSIDERS THE INVESTIGATION CLOSED.

Description of Event or Problem · 1

IT IS REPORTED THAT THE PT WAS REVISED IN (B)(6) 2011 DUE TO PAIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
32554 NEXGEN GSF LPS-FLEX FEMORAL COMPONENT JWH ZIMMER INC 60788903

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention