Report Number
1644408-2012-00622
Event Type
Other
Date Received
November 19, 2012
Date of Event
November 12, 2012
Report Date
November 12, 2012
Manufacturer
ENCORE MEDICAL, L.P.
Product Code
JWH
PMA / PMN Number
K933539
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REASON FOR THIS REVISION SURGERY WAS TO REMEDY A WORN POLY INSERT AFTER 14.5 YEARS OF PATIENT USE. THE OUTCOMES ATTRIBUTED TO THIS EVENT ARE REQUIRED INTERVENTION TO PREVENT PERMANENT IMPAIRMENT/DAMAGE. THE HEALTHCARE PROFESSIONAL INDICATED A SIGNIFICANT ADVERSE EVENT TO THE PATIENT. THERE IS NO INFORMATION IN THIS COMPLAINT ABOUT ANY PATIENT INJURIES, ACTIVITIES, ACCIDENTS, OR MEDICAL CONTRAINDICATIONS THAT MAY HAVE CONTRIBUTED TO THE REVISION SURGERY. THERE WAS NO DELAY IN SURGERY AND ANOTHER SUITABLE DEVICE WAS AVAILABLE FOR USE. THE REVISION SURGERY WAS COMPLETED AS INTENDED. THE DEVICE WAS NOT RETURNED TO DJO SURGICAL FOR EXAMINATION. A REVIEW OF THE DEVICE HISTORY RECORDS SHOWED NO NON-CONFORMING MATERIAL REPORTS ASSOCIATED WITH THIS PRODUCT. A REVIEW OF THE PRODUCT COMPLAINT REPORT HISTORY SHOWS THIS IS THE SECOND COMPLAINT FOR THIS PART NUMBER, ONE OTHER FOR A CRACKED/BROKEN DEVICE. THIS IS THE FIRST COMPLAINT FOR THIS LOT NUMBER. THE ROOT CAUSE FOR THE WORN POLY INSERT WAS NOT DETERMINED WITH CONFIDENCE. THERE IS NO INFORMATION REPORTED THAT SHOWED A MATERIAL, DESIGN, OR MANUFACTURING PROBLEM WITH THE PRODUCT. THERE ARE NO INDICATIONS OF A PRODUCT OR PROCESS ISSUE AFFECTING IMPLANT SAFETY OR EFFECTIVENESS.

Description of Event or Problem · 1

REVISION SURGERY- THE PT HAD POLYETHYLENE WEAR ON SERIES (B)(4) POSTERIOR STABILIZED INSERT. THE ORIGINAL SURGERY WAS PERFORMED (B)(6) AGO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FOUNDATION KNEE POSTERIOR STABILIZED INSERT, 11MM JWH ENCORE MEDICAL, L.P. 147191