FDA Adverse Event Injury Summary report: N

UNKNOWN SEGMENTAL KNEE FEMORAL COMPONENT

MDR report key: 2822423 · Received November 2, 2012

Report

Report Number
1822565-2012-02267
Event Type
Injury
Date Received
November 2, 2012
Date of Event
October 4, 2012
Report Date
October 4, 2012
Manufacturer
ZIMMER, INC.
Product Code
JWH
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: SURGICAL NOTES WERE NOT PROVIDED. X-RAYS WERE NOT PROVIDED; IT IS UNKNOWN WHETHER THE COMPONENTS WERE IMPLANTED WITH THE CORRECT FIT AND ORIENTATION AS PER THE SURGICAL TECHNIQUE. PATIENT FACTORS THAT MAY AFFECT THE PERFORMANCE OF THE COMPONENTS SUCH AS AGE, BONE QUALITY, HEIGHT/WEIGHT, ACTIVITY LEVEL, TYPE OF ACTIVITY (LOW IMPACT VS. HIGH IMPACT) ARE UNKNOWN. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED WITH THE INFORMATION PROVIDED. EVALUATION: REVIEW OF THE DEVICE HISTORY RECORDS WAS NOT POSSIBLE AS THE PRODUCT AND/OR LOT NUMBERS REQUIRED FOR RETRIEVAL WERE UNAVAILABLE. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE AVAILABLE INFORMATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL SUBSTANTIVE INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. ZIMMER, INC. CONSIDERS THE INVESTIGATION CLOSED.

Description of Event or Problem · 1

IT IS REPORTED THAT THE PATIENT UNDERWENT AN ADDITIONAL PROCEDURE DUE TO MIGRATION OF THE FEMORAL HINGE PIN. IT WAS ALSO REPORTED THAT DURING ORIGINAL IMPLANTATION, THE SECONDARY LOCKING SET SCREW WAS NOT INSTALLED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNKNOWN SEGMENTAL KNEE FEMORAL COMPONENT KNEE PROSTHESIS JWH ZIMMER, INC.

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention