PFC*SIGMA/OV/DOME PAT 3PEG,38
Report
- Report Number
- 1818910-2012-23294
- Event Type
- Injury
- Date Received
- October 11, 2012
- Date of Event
- October 3, 2012
- Report Date
- October 3, 2012
- Product Code
- JWH
- PMA / PMN Number
- K961685
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE AS THE LOT CODES REQUIRED WERE NOT PROVIDED. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A; REV. C. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT WITH THE INFORMATION AVAILABLE. BASED ON THE INABILITY TO IDENTIFY A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ANY ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.
PATIENT WAS REVISED TO ADDRESS POLY WEAR OF THE INSERT AND PATELLA AND LOOSENESS IN FLEXION AND EXTENSION. IT WAS REPORTED THAT THE PATELLA WAS LOOSE AT THE CEMENT/IMPLANT INTERFACE, WITH THE LATERAL PEG WAS SHEARED OFF THE PATELLA, AND THE TWO MEDIAL PEGS WERE BENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PFC*SIGMA/OV/DOME PAT 3PEG,38 | PATELLA PROSTHESIS | JWH | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Required Intervention |