FDA Adverse Event Injury Summary report: N

EVOLIS TOTAL KNEE SYSTEM

MDR report key: 2613841 · Received February 19, 2010

Report

Report Number
3005180920-2010-00003
Event Type
Injury
Date Received
February 19, 2010
Date of Event
November 1, 2009
Report Date
November 26, 2009
Manufacturer
MEDACTA INTERNATIONAL, SA
Product Code
JWH
PMA / PMN Number
K081023
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

MFG RECORDS AND QUALITY DOCUMENTS WERE REVIEWED. THE MFG AND QUALITY DOCUMENT REVIEW CONFIRMED THAT ALL POSTERO STABILIZED CEMENTED FEMURS RELEASED TO THE FIELD OF LOT NUMBER 021206 (8 MFG AND 4 SENT TO (B)(6), 2 TO (B)(6), 2 TO (B)(6)) CONFORM TO THE SPECIFICATION VALID AT THE TIME OF MFR, INCLUDING DIMENSIONAL AND SHAPE VALUES. THE MFG AND QUALITY DOCUMENT REVIEW CONFIRMED THAT ALL POSTERO STABILIZED INSERTS RELEASED TO THE FIELD OF LOT NUMBER 021153 (5 MFG AND 2 SENT TO (B)(6), 2 TO (B)(6), 1 TO (B)(6)) CONFORM TO THE SPECIFICATION VALID AT THE TIME OF MFR, INCLUDING DIMENSIONAL SHAPE VALUES. THE MFG AND QUALITY DOCUMENT REVIEW CONFIRMED THAT ALL TIBIAL BASEPLATES RELEASED TO THE FIELD OF LOT NUMBER 031571 (30 MFG AND 2 SENT TO (B)(6), 4 TO (B)(6), 1 TO (B)(6), 23 TO (B)(6)) CONFORM TO THE SPECIFICATION VALID AT THE TIME OF MFR, INCLUDING DIMENSIONAL AND SHAPE VALUES. THE DEVICE HAS BEEN RETURNED AND AN INVESTIGATION INTO THE ROOT CAUSE AND POSSIBLE CORRECTIVE ACTION IS IN PROGRESS BUT NOT COMPLETE AT THIS TIME.

Description of Event or Problem · 1

A TOTAL KNEE REVISION SURGERY WAS PERFORMED APPROX. FIVE YEARS FROM THE PRIMARY SURGERY. THE REVISION WAS REQUIRED DUE TO BREAKAGE OF THE POSTERO-STABILIZATION PEG OF THE INSERT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EVOLIS TOTAL KNEE SYSTEM POSTERO-STABILIZED INSERT 9MM SIZE 5 JWH MEDACTA INTERNATIONAL, SA NA 021153

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention PS CEMENTED FEMUR SIZE 5 RIGHT: (B)(4),| LOT 021206| CEMENTED TIBIAL BASEPLATE SIZE 5: (B)(4),| LOT 031571