PHYSICA TT
Report
- Report Number
- 3008021110-2026-00227
- Event Type
- Injury
- Date Received
- May 21, 2026
- Date of Event
- December 4, 2025
- Report Date
- May 19, 2026
- Manufacturer
- LIMACORPORATE S.P.A.
- Product Code
- JWH
- UDI-DI
- 08033390169912
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE CHECK OF DHRS OF THE INVOLVED LOT NUMBER 2211436 DID NOT HIGHLIGHT ANY PRE-EXISTING ANOMALIES ON THE ITEMS MANUFACTURED WITH THIS LOT NUMBER. THIS IS THE FIRST AND ONLY COMPLAINT RECEIVED ON THIS LOT NUMBER. A FINAL REPORT WILL BE SUBMITTED AFTER THE INVESTIGATION IS COMPLETED.
ON (B)(6) 2025 REVISION OF KNEE PROTHESIS DUE TO SINTERING OF PHYSICA TT FIX. TIBIAL TRAY #5 (PRODUCT CODE: 6521.14.050, LOT: 2211436 - STER: (B)(4). DURING REVISION THE TOTAL PRIMARY KNEE HAS BEEN CHANGED TO A ROTATING HINGED KNEE. PREVIOUS SURGERY WAS PERFORMED ON (B)(6) 2023. DURING ONE YEAR-CONTROL ONLY MINOR RADIOLOGIC SIGNS OF SINTERING SHOWED UP. INCREASING SINTERING WITHIN THE SECOND YEAR OF POST-OP. PATIENT DID NOT REPORT ANY TRAUMA. PRODUCT CODES AND LOT NUMBERS OF THE OTHER DEVICES IMPLANTED ON THE FIRST SURGERY ON 2023: PHYSICA TT FIX. TIBIAL TRAY #5 (PRODUCT CODE: 6521.14.050, LOT: 2211436 - STER: (B)(4), PS FEMORAL COMPONENT #5 LEFT (PRODUCT CODE: 6515.10.550, LOT: 2222424 - STER: (B)(4), PHYSICA PS TIBIAL LINER #5 (PRODUCT CODE: 6535.50.511, LOT: 22BC069 - STER: (B)(4), PHYSICA PS - FEMORAL PEG (PRODUCT CODE: 6515.09.900, LOT: 2313310 - STER: (B)(4). PATIENT: FEMALE, 75 YEARS OLD, APPROXIMATE HEIGHT 1.65(M), APPROXIMATE WEIGHT 58 (KG), ACTIVE LEVEL, NO SMOKING. EVENT OCCURRED IN AUSTRIA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 129115 | PHYSICA TT | PHYSICA TT - TI6AL4V EBM - FIXED TIBIAL PLATE #5 UNCEMENTED | JWH | LIMACORPORATE S.P.A. | 6521.14.050 | 2211436 | 08033390169912 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |