FDA Adverse Event Malfunction Summary report: N

BD PYXIS¿ ANESTHESIA STATION ES

MDR report key: 25233882 · Received May 20, 2026

Report

Report Number
2016493-2026-30627
Event Type
Malfunction
Date Received
May 20, 2026
Date of Event
April 24, 2026
Report Date
April 29, 2026
Manufacturer
CAREFUSION 303, INC.
Product Code
BRY
UDI-DI
10885403533242
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE COMPLAINT HISTORY FOR SN (B)(6)WAS PERFORMED IN SALESFORCE WHICH DID NOT LOCATE SIMILAR COMPLAINT(S) WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD FOR SN (B)(6)WAS PERFORMED FROM THE DATE OF MANUFACTURE, 05-NOV-2022 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THE STATION DATABASE WAS BLOATED TO 9 GB. A TECHNICAL SUPPORT SPECIALIST (TSS) DIALED INTO THE STATION AND PERFORMED A MANUAL DATABASE SHRINK USING STRUCTURED QUERY LANGUAGE (SQL), WHICH REDUCED THE DATABASE SIZE TO 5.9 GB. IN ANOTHER CASE TSS PERFORMED A FULL SYNC WITH DATABASE DROP AND REPAIR, REDUCING THE ISSUE, AND DATA WAS CONFIRMED AS CURRENT WITH NO PENDING TRANSACTIONS. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE TECHNICAL SUPPORT SPECIALIST TROUBLESHOT THE DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN USING THE BD PYXIS¿ ANESTHESIA STATION ES, CURRENT PATIENT PROFILES WERE NOT VISIBLE ON THE DEVICE, AND ONLY TEMPORARY PATIENTS APPEARED. NO ERROR MESSAGES WERE DISPLAYED. PATIENTS WERE VISIBLE IN GLOBAL PATIENT SEARCH (EPHI). THE CUSTOMER REBOOTED THE DEVICE; HOWEVER, THE ISSUE PERSISTED. THE CUSTOMER STATED THAT THERE WAS A DELAY IN PATIENT CARE. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
560312 BD PYXIS¿ ANESTHESIA STATION ES AUTOMATED DISPENSING CABINET BRY CAREFUSION 303, INC. 500400001500 10885403533242

Patients

Seq Age Sex Outcome Treatment
1