FDA Adverse Event Injury Summary report: N

GMK SPHERE TOTAL KNEE SYSTEM

MDR report key: 25148965 · Received May 12, 2026

Report

Report Number
3005180920-2026-00432
Event Type
Injury
Date Received
May 12, 2026
Date of Event
April 21, 2026
Report Date
May 12, 2026
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JWH
UDI-DI
07630971261518
PMA / PMN Number
K202022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 28 ARPIL 2026 GMK-SPHERIKA 02.12.E0310FL GMK-SPHERE TIBIAL INSERT E-CROSS - FLEX 3L - 10MM (K202022) LOT 2431924: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 04-FEB-2025. EXPIRATION DATE: 20-JAN-2030. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH ONE SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. GMK-SPHERIKA 02.12.KA13L GMK SPHERIKA FEMORAL COMPONENT S3+L CEMENTED (K211004) LOT 2347599: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 22-APR-2024. EXPIRATION DATE: 04-APR-2029. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. GMK-SPHERIKA 02.12.T4I3L GMK-SPHERE TIBIAL COMPONENT CEMENTED T4I3L (K121416) LOT 2242814: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 17-FEB-2023. EXPIRATION DATE: 01-FEB-2028. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ROOT CAUSE: INFECTION IS A KNOWN POSSIBLE COMPLICATION OF ANY SURGERY. ALTHOUGH NO ROOT CAUSE CAN BE ESTABLISHED, THERE IS NO INDICATION THAT ANY ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AND THE DOCUMENT REVIEW DOES NOT INDICATE ANY POTENTIAL MANUFACTURING RELATED CAUSE.

Description of Event or Problem · 0

AT ABOUT 11 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE GMK-SPHERIKA COMPONENTS TO GMK-HINGE. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
376982 GMK SPHERE TOTAL KNEE SYSTEM GMK-SPHERE TIBIAL INSERT E-CROSS - FLEX 3L - 10MM JWH MEDACTA INTERNATIONAL SA 02.12.E0310FL 2431924 07630971261518

Patients

Seq Age Sex Outcome Treatment
1