FDA Adverse Event Malfunction Summary report: N

BD PYXIS¿ MEDSTATION¿ ES

MDR report key: 25135316 · Received May 11, 2026

Report

Report Number
2016493-2026-27732
Event Type
Malfunction
Date Received
May 11, 2026
Date of Event
April 13, 2026
Report Date
April 15, 2026
Manufacturer
CAREFUSION 303, INC.
Product Code
BRY
UDI-DI
10885403533228
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE COMPLAINT HISTORY FOR SN (B)(6) WAS PERFORMED IN SALESFORCE WHICH DID NOT LOCATE SIMILAR COMPLAINT(S) WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD FOR SN (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE, 23-MAR-2021 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THERE WAS NO FAILED DRAWER. A FIELD SERVICE ENGINEER (FSE) RESEATED AND CHECKED THE 2.1 CABLES FOR BOTH THE MAIN AND AUXILIARY DRAWERS, INSPECTED FOR MISSED OR DEFECTIVE SLIDE BUMPERS, AND THE CUSTOMER TESTED THE UNIT WITH NO ISSUES. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE FIELD SERVICE ENGINEER INSPECTED THE DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN USING THE BD PYXIS¿ MEDSTATION¿ ES, THE CUBIE DRAWER 2.2 POCKET E1 LID WAS NOT OPENING, WHICH LOCATED ON 2KM. THE CUSTOMER ATTEMPTED TO REBOOT THE STATION, BUT THE ISSUE PERSISTED. THE CUSTOMER STATED THAT THIS MALFUNCTION OCCURRED WHILE DISPENSING THE MEDICATION AND THE USER MANAGED TO GET THE MEDICATION FROM ANOTHER STATION, WHICH CAUSED A DELAY TO THE PATIENT CARE. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
220217 BD PYXIS¿ MEDSTATION¿ ES AUTOMATED DISPENSING CABINET BRY CAREFUSION 303, INC. 500001002500 10885403533228

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown