BD PYXIS¿ MEDSTATION¿ ES
Report
- Report Number
- 2016493-2025-141608
- Event Type
- Malfunction
- Date Received
- December 9, 2025
- Date of Event
- November 21, 2025
- Report Date
- November 27, 2025
- Manufacturer
- CAREFUSION 303, INC.
- Product Code
- BRY
- UDI-DI
- 10885403533228
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- PHARMACIST
- Health Professional
- Yes
Narratives
A REVIEW OF THE COMPLAINT HISTORY FOR SN (B)(6) WAS PERFORMED IN SALESFORCE WHICH DID NOT LOCATE SIMILAR COMPLAINTS WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD FOR SN (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE, 09-AUG-2017 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT USERS WERE UNABLE TO LOG IN. THE TECHNICAL SUPPORT SPECIALIST (TSS) DETERMINED DEVICE HAD A DATABASE LOG FILE OCCUPYING 72 GB, LEAVING ONLY 200 MB OF FREE SPACE ON THE D DRIVE. THE DSCLIENTOLTP DATABASE WAS CHECKED IN SQL SERVER MANAGEMENT STUDIO, AND ITS RECOVERY MODE WAS SET TO FULL. THE RECOVERY MODE WAS CHANGED TO SIMPLE, AND THE DATABASE WAS SHRUNK, REDUCING THE LOG FILE TO APPROXIMATELY 500 MB. THIS FREED UP 74 GB OF SPACE ON THE D DRIVE. AFTER THESE ACTIONS, THE CUSTOMER TESTED LOGIN SUCCESSFULLY, AND THE CONFIRMED THE ISSUE WAS RESOLVED. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE TECHNICAL SUPPORT SPECIALIST TROUBLESHOT THE DEVICE.
IT WAS REPORTED THAT WHEN USING THE BD PYXIS¿ MEDSTATION¿ ES HAD DEVICE ON CRITICAL OVERRIDE AND USERS UNABLE TO LOG IN. THE CUSTOMER STATED THAT THIS MALFUNCTION OCCURRED WHILE DISPENSING THE MEDICATION AND CAUSED A DELAY TO THE PATIENT CARE. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2813706 | BD PYXIS¿ MEDSTATION¿ ES | AUTOMATED DISPENSING CABINET | BRY | CAREFUSION 303, INC. | 500001002500 | 10885403533228 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |