FDA Adverse Event Malfunction Summary report: N

BD PYXIS¿ ANESTHESIA STATION ES

MDR report key: 23603150 · Received November 20, 2025

Report

Report Number
2016493-2025-136546
Event Type
Malfunction
Date Received
November 20, 2025
Date of Event
October 23, 2025
Report Date
November 5, 2025
Manufacturer
CAREFUSION 303, INC.
Product Code
BRY
UDI-DI
10885403533242
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
PHARMACIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE COMPLAINT HISTORY FOR (B)(6) WAS PERFORMED IN SALESFORCE WHICH DID NOT LOCATE SIMILAR COMPLAINT(S) WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD FOR (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE, 14-DEC-2023 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THE DEVICE HAD AN ISSUE WHERE THE USER WAS UNABLE TO INPUT WASTE MEDICATION. A TECHNICAL SUPPORT SPECIALIST CHECKED THE LOGS, PERFORMED HOTFIX AS PER KNOWLEDGE ARTICLE (KA) (B)(4)¿UNABLE TO ENTER WASTE AMOUNT ON PAS ES; NUMBER KEYS UNRESPONSIVE¿, REBOOTED THE STATION, REPAIRED APPLICATION AND VERIFIED DATABASE RECOVERY MODE TO RESOLVE THE ISSUE. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE TECHNICAL SUPPORT SPECIALIST TROUBLESHOT THE DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN USING THE BD PYXIS¿ ANESTHESIA STATION ES, STATION WAS UNABLE TO INPUT WASTE MEDICATION. THE CUSTOMER STATED THAT THIS MALFUNCTION OCCURRED WHILE DISPENSING THE MEDICATION WHICH CAUSED A DELAY TO THE PATIENT CARE. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2529061 BD PYXIS¿ ANESTHESIA STATION ES AUTOMATED DISPENSING CABINET BRY CAREFUSION 303, INC. 500400001500 10885403533242

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown