Narratives
UPON INVESTIGATION OF THE ACTUAL DEVICE IN THIS INCIDENT. IT WAS DETERMINED THAT THE ORDERS WERE NOT CROSSING OVER TO STATION. A TECHNICAL SUPPORT SPECIALIST (TSS) CHECKED THE STATION AND FOUND NO ALERTS FOR PATIENTS OR ORDERS. TSS ASKED CUSTOMER IF USER HAD ANY FEEDBACK FROM HOSPITAL INFORMATION TECHNOLOGY TEAM (HIT) AND VERIFIED SSMS FOR SITE STATUS¿NO DISCONNECTIONS OR DELAYS. TSS REVIEWED PATIENT EXAMPLES IN EPHI AND CONFIRMED TWO DISCONTINUED ORDERS ON THE SERVER, WHILE TWO NEWER ORDERS FOR THE SAME PATIENT WERE MISSING. TSS RESTARTED SEVERAL SERVICES AND ESCALATED TO IEST TO VERIFY MESSAGES IN CAREFUSION COORDINATION ENGINE (CCE). THE TSS DIALED INTO CCE, FOUND TWO SOLUTIONS, AND RESTARTED THE CCE-SERVICE FOR MED. WHILE ON THE CALL, USAGE DROPPED, SUGGESTING HIT WAS MAKING PROGRESS, BUT AFTER DISCONNECTING, TSS WAS UNABLE TO PING. THE TSS RESTARTED THE CCE-SERVICE AGAIN ON THE MED SOLUTION; STILL NO INBOUND ADT/RDE. TSS AFTER REFRESHING THE MESSAGE TRACE, ACTIVE DIRECTORY AND RDE MESSAGES BEGAN CROSSING FROM RX CONNECT. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE TECHNICAL SUPPORT SPECIALIST TROUBLESHOT THE DEVICE. D4 & H4: SERIAL NUMBER, CATALOG, MODEL, UNIQUE DEVICE IDENTIFIER AND MANUFACTURER DATE NOT AVAILABLE.
IT WAS REPORTED THAT WHEN USING THE BD PYXIS¿ ES SERVER ORDERS WERE NOT CROSSING OVER TO STATION. THE CUSTOMER STATED THAT THERE WAS A DELAY IN PATIENT CARE. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED DUE TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1854107 | BD PYXIS¿ ES SERVER | AUTOMATED DISPENSING CABINET | BRY | CAREFUSION 303, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | SN: (B)(6) LOCATION: (B)(6) |