BD PYXIS¿ MEDSTATION¿ ES
Report
- Report Number
- 2016493-2025-123916
- Event Type
- Malfunction
- Date Received
- October 17, 2025
- Date of Event
- May 2, 2023
- Report Date
- September 26, 2025
- Manufacturer
- CAREFUSION 303, INC.
- Product Code
- BRY
- UDI-DI
- 10885403512667
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS MDR IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW OF RECORDS DATING JANUARY 2022-DECEMBER 5, 2024, UNDER CAPA 10308384. THE LATE SUBMISSION OF THIS REPORT IS JUSTIFIED BY THE THOROUGH AND DETAILED NATURE OF THE RETROSPECTIVE REVIEW PROCESS. THIS PROCESS WAS ESSENTIAL TO ACCURATELY CAPTURE ALL RELEVANT DATA AND ENSURE THE INTEGRITY OF OUR REPORTING. WE HAVE INCLUDED THE CAPA REFERENCE FOR THE RETROSPECTIVE REVIEW IN THE ADDITIONAL MANUFACTURER NARRATIVE SECTION OF THE FDA FORM 3500A, AS RECOMMENDED. A REVIEW OF THE COMPLAINT HISTORY FOR SN (B)(6) WAS PERFORMED IN SALESFORCE WHICH DID LOCATE 1 SIMILAR COMPLAINT WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. ¿ CASE: 3191666 ¿ D-ES- STN R2E-2E01 TRANSACTIONS PRINTER NOT PRINTING AFTER RESTART A REVIEW OF THE DEVICE HISTORY RECORD FOR SN (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE, 13-MAR-2014 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THE DEVICE HAD NO RECORDS ON MEDICATION IN PYXIS AND THE CUBIE WAS STUCK WHEN NURSE WAS REMOVING THE MEDICATION. A TECHNICAL SUPPORT SPECIALIST CHECKED THE LOGS AND FOUND THAT A TIMEOUT EVENT OCCURRED, WHICH PREVENTED THE TRANSACTION FROM CAPTURED AND WAS NOT RECORDED BY THE SERVER AND DEVICE DATABASE. THE SYSTEM ASSESSED BY A TECHNICAL SUPPORT SPECIALIST.
IT WAS REPORTED BY THE CUSTOMER THAT A BD PYXIS MEDSTATION ES HAD NO RECORDS ON MEDICATION IN PYXIS AND THE CUBIE WAS STUCK WHEN NURSE WAS REMOVING THE MEDICATION. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 299098 | BD PYXIS¿ MEDSTATION¿ ES | AUTOMATED DISPENSING CABINET | BRY | CAREFUSION 303, INC. | 323 | 10885403512667 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |