FDA Adverse Event Injury Summary report: N

LOGIC TIBIA PS MOD INSRT SZ 2.5 9MM

MDR report key: 23235707 · Received October 7, 2025

Report

Report Number
1038671-2025-03056
Event Type
Injury
Date Received
October 7, 2025
Date of Event
September 15, 2025
Report Date
October 7, 2025
Manufacturer
EXACTECH, INC.
Product Code
JWH
PMA / PMN Number
K160484
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CONCOMITANTS: 02-010-01-0225 - LOGIC FEMORAL PS CEM LEFT SZ 2.5: (B)(6). 02-012-35-2509 - LOGIC TIBIA PS MOD INSRT SZ 2.5 9MM: (B)(6). 02-012-45-2525 - LGC TIBIAL FIT TRAY CEM SZ 2.5F / 2.5T: (B)(6). 200-02-32 - THREE PEG PATELLA 32MM: (B)(6). RECALL NUMBER: Z-0021-2022. THE REVISION WAS LIKELY THE RESULT OF PROSTHESIS WEAR OVER THE COURSE OF OVER 12 YEARS OF IMPLANTATION. POSSIBLE CAUSES FOR POLYETHYLENE DAMAGE INCLUDE HIGH CONTACT STRESSES DURING KNEE FLEXION, UNEQUAL JOINT FORCE DISTRIBUTION BETWEEN THE MEDIAL AND LATERAL SIDES, AND INCLUSION OF THE POLYETHYLENE IN THE PACKAGING RECALL, OR ANY COMBINATION OF THESE POSSIBILITIES. POTENTIAL CONTRIBUTIONS OF USER AND PATIENT-RELATED CONSIDERATIONS TO THE EVENT COULD NOT BE ASSESSED AS THE DEVICES WERE NOT AVAILABLE FOR EVALUATION AND RELEVANT CLINICAL NOTES WERE NOT PROVIDED. IF ANY FURTHER INFORMATION IS OBTAINED THAT WOULD CHANGE OR ALTER ANY INFORMATION PROVIDED, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL LEFT TOTAL KNEE ARTHROPLASTY. SUBSEQUENTLY, THE PATIENT COMPLAINED OF PAIN, STIFFNESS AND DISSATISFACTION WITH THEIR KNEE REPLACEMENT. AS A RESULT, THE PATIENT UNDERWENT A LEFT KNEE REVISION APPROXIMATELY 12 YEARS AND 6 MONTHS AFTER INITIAL IMPLANTATION. PROVIDED IMAGES OF THE TIBIAL INSERT SHOW SIGNS OF PROSTHESIS WEAR. NO FURTHER PATIENT IMPACT REPORTED. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1869276 LOGIC TIBIA PS MOD INSRT SZ 2.5 9MM PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL JWH EXACTECH, INC.

Patients

Seq Age Sex Outcome Treatment
1 83 YR Female Required Intervention| H SEE H11