BD PYXIS¿ MEDBANK TOWER
Report
- Report Number
- 2016493-2025-108940
- Event Type
- Malfunction
- Date Received
- August 28, 2025
- Date of Event
- August 4, 2025
- Report Date
- September 2, 2025
- Manufacturer
- CAREFUSION 303, INC.
- Product Code
- BRY
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
ADDITIONAL INFORMATION: SECTION H IMDRF ANNEX CODES. A REVIEW OF THE COMPLAINT HISTORY FOR SN (B)(6) WAS PERFORMED IN SALESFORCE WHICH DID NOT LOCATE SIMILAR COMPLAINT(S) WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD FOR SN (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE, 02-FEB-2023 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THE USER WAS UNABLE TO ISSUE MEDICATION. A TECHNICAL SUPPORT SPECIALIST WALKED THE CUSTOMER THROUGH CYCLE COUNTING THE BIN TO UPDATE THE QUANTITY ON HAND (QOH). THE SPECIALIST ADVISED THE CUSTOMER TO CONTACT THE PHARMACY TO HAVE THE RESTOCK SENT TO THE FACILITY AS SOON AS POSSIBLE. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE TECHNICAL SUPPORT SPECIALIST VERIFIED THE DEVICE.
IT WAS REPORTED BY THE CUSTOMER THAT A BD PYXIS¿ MEDBANK TOWER UNABLE TO ISSUE. THE CUSTOMER STATED THAT THIS MALFUNCTION CAUSED A DELAY IN DISPENSING MEDICATION TO PATIENTS. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS EVENT.
IT WAS REPORTED BY THE CUSTOMER THAT A BD PYXIS¿ MEDBANK TOWER UNABLE TO ISSUE. THE CUSTOMER STATED THAT THIS MALFUNCTION CAUSED A DELAY IN DISPENSING MEDICATION TO PATIENTS. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1419367 | BD PYXIS¿ MEDBANK TOWER | AUTOMATED DISPENSING CABINET | BRY | CAREFUSION 303, INC. | 500004500000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |