BD PYXIS¿ MEDSTATION¿ ES
Report
- Report Number
- 2016493-2025-102913
- Event Type
- Malfunction
- Date Received
- August 6, 2025
- Date of Event
- July 9, 2025
- Report Date
- August 7, 2025
- Manufacturer
- CAREFUSION 303, INC.
- Product Code
- BRY
- UDI-DI
- 10885403512667
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
ADDITIONAL INFORMATION: SECTION H IMDRF ANNEX CODES. A REVIEW OF THE COMPLAINT HISTORY FOR SN (B)(6) WAS PERFORMED IN SALESFORCE WHICH DID NOT LOCATE SIMILAR COMPLAINT(S) WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD FOR SN (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE, 24-APR-2023 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THE DRAWER 1 WAS NOT OPENING. A FIELD SERVICE ENGINEER POWERED DOWN THE STATION AND DISASSEMBLED THE DRAWER PER FSG SECTION 3.11.3, REPLACED THE ROW TRAY, RESEATED THE RETRACTOR BAND AND ROW BOARD CABLES AT BOTH THE DRAWER AND MODULE CONTROLLER, REASSEMBLED THE DRAWER, POWERED ON THE DEVICE, AND VERIFIED DRAWER FUNCTIONALITY THROUGH DIAGNOSTICS. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE FIELD SERVICE ENGINEER REPAIRED THE DEVICE.
IT WAS REPORTED BY THE CUSTOMER THAT A BD PYXIS¿ MEDSTATION¿ ES DRAWER WAS NOT OPENING. THE CUSTOMER STATED THERE WAS A DELAY TO THE PATIENT'S WORKFLOW. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS EVENT.
IT WAS REPORTED BY THE CUSTOMER THAT A BD PYXIS¿ MEDSTATION¿ ES DRAWER WAS NOT OPENING. THE CUSTOMER STATED THERE WAS A DELAY TO THE PATIENT'S WORKFLOW. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1725180 | BD PYXIS¿ MEDSTATION¿ ES | AUTOMATED DISPENSING CABINET | BRY | CAREFUSION 303, INC. | 500001002500 | 10885403512667 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |