FDA Adverse Event Malfunction Summary report: N

BD PYXIS¿ MEDSTATION¿ ES

MDR report key: 22688426 · Received August 1, 2025

Report

Report Number
2016493-2025-102074
Event Type
Malfunction
Date Received
August 1, 2025
Date of Event
January 21, 2024
Report Date
July 31, 2025
Manufacturer
CAREFUSION 303, INC.
Product Code
BRY
UDI-DI
10885403512667
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS MDR IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW OF RECORDS DATING JANUARY 2022-DECEMBER 5, 2024, UNDER CAPA 10308384. THE LATE SUBMISSION OF THIS REPORT IS JUSTIFIED BY THE THOROUGH AND DETAILED NATURE OF THE RETROSPECTIVE REVIEW PROCESS. THIS PROCESS WAS ESSENTIAL TO ACCURATELY CAPTURE ALL RELEVANT DATA AND ENSURE THE INTEGRITY OF OUR REPORTING. WE HAVE INCLUDED THE CAPA REFERENCE FOR THE RETROSPECTIVE REVIEW IN THE ADDITIONAL MANUFACTURER NARRATIVE SECTION OF THE FDA FORM 3500A, AS RECOMMENDED. A REVIEW OF THE COMPLAINT HISTORY FOR SN (B)(6) WAS PERFORMED IN SALESFORCE WHICH DID NOT LOCATE SIMILAR COMPLAINTS WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD FOR SN (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE 18-JUL-2022 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT PRN ORDERS NOT REACTIVATING WHEN PATIENT RETURNS FROM LOA. TECHNICAL SUPPORT SPECIALIST HAD STATED THAT THE LOA MESSAGES WERE FOUND INCOMPATIBLE WITH ES, AND THE ISSUE WAS ESCALATED TO ADT SUPPORT. NO FURTHER INFORMATION WAS AVAILABLE.

Description of Event or Problem · 0

IT WAS REPORTED BY THE CUSTOMER THAT BD PYXIS¿ MEDSTATION¿ ES HAD PRN ORDERS NOT REACTIVATING WHEN PATIENT RETURNS FROM LOA. THE CUSTOMER STATED THAT THIS WAS A RECURRING ISSUE AT THEIR FACILITY AND COULD POSE A PATIENT SAFETY RISK, AS PRN MEDS ARE OFTEN NEEDED FOR EMERGENT SITUATIONS. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290218 BD PYXIS¿ MEDSTATION¿ ES AUTOMATED DISPENSING CABINET BRY CAREFUSION 303, INC. 323 10885403512667

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown