FDA Adverse Event Malfunction Summary report: N

BD PYXIS¿ MEDSTATION¿ ES

MDR report key: 22459124 · Received July 9, 2025

Report

Report Number
2016493-2025-94867
Event Type
Malfunction
Date Received
July 9, 2025
Date of Event
June 13, 2025
Report Date
July 8, 2025
Manufacturer
CAREFUSION 303, INC.
Product Code
BRY
UDI-DI
10885403512667
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE COMPLAINT HISTORY FOR SN (B)(6) WAS PERFORMED IN SALESFORCE, WHICH DID NOT LOCATE SIMILAR COMPLAINT(S) WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD FOR SN (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE, 12-AUG-2015 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT DRAWER 3 ON MAIN WAS NOT STAYING CLOSED. THE YELLOW PIECE HOUSING THE RED EMERGENCY RELEASE WAS DETACHED, AND THE EMERGENCY RELEASE ITSELF WAS BROKEN. BROKEN SCREWS WERE PRESENT, AND PHARMACY STAFF HAD DAMAGED QBS WHILE ATTEMPTING TO ACCESS MEDICATION. A FIELD SERVICE ENGINEER REATTACHED THE YELLOW HOUSING, INSTALLED A NEW EMERGENCY RELEASE, REBOOTED THE SYSTEM, AND VERIFIED FUNCTIONALITY USING THE HARDWARE TEST APPLICATION. DEBRIS WAS CLEARED FROM THE ROW BOARDS, BROKEN QBS WERE REMOVED, AND ESCORT ACCESS TO THE PYXIS WAS GRANTED. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE FIELD SERVICE ENGINEER REPAIRED THE DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED BY THE CUSTOMER THAT A BD PYXIS¿ MEDSTATION¿ ES DRAWER WAS BROKEN AND WOULD NOT STAY CLOSED. THE CUSTOMER STATED THAT THIS MALFUNCTION CAUSED A DELAY IN DISPENSING MEDICATION TO PATIENTS. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
701765 BD PYXIS¿ MEDSTATION¿ ES AUTOMATED DISPENSING CABINET BRY CAREFUSION 303, INC. 323 10885403512667

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown