BD PYXIS¿ MEDSTATION¿ ES
Report
- Report Number
- 2016493-2025-93493
- Event Type
- Malfunction
- Date Received
- July 5, 2025
- Date of Event
- December 8, 2023
- Report Date
- June 13, 2025
- Manufacturer
- CAREFUSION 303, INC.
- Product Code
- BRY
- UDI-DI
- 10885403512667
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS MDR IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW OF RECORDS DATING (B)(6) 2022-(B)(6) 2024, UNDER CAPA 10308384. THE LATE SUBMISSION OF THIS REPORT IS JUSTIFIED BY THE THOROUGH AND DETAILED NATURE OF THE RETROSPECTIVE REVIEW PROCESS. THIS PROCESS WAS ESSENTIAL TO ACCURATELY CAPTURE ALL RELEVANT DATA AND ENSURE THE INTEGRITY OF OUR REPORTING. WE HAVE INCLUDED THE CAPA REFERENCE FOR THE RETROSPECTIVE REVIEW IN THE ADDITIONAL MANUFACTURER NARRATIVE SECTION OF THE FDA FORM 3500A, AS RECOMMENDED. A REVIEW OF THE COMPLAINT HISTORY FOR SN (B)(6) WAS PERFORMED IN SALESFORCE WHICH DID NOT LOCATE SIMILAR COMPLAINTS WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD FOR SN (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE, 27-MAR-2023 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THE DEVICE KEPT GLITCHING. A TECHNICAL SUPPORT SPECIALIST CONFIRMED ALL TRANSACTIONS UPLOADED AND REPAIRED MED APPLICATION AND FULL SYNCED STATION TO RESOLVE THE ISSUE. THE SYSTEM FUNCTIONED AS INTENDED AFTER TROUBLESHOT BY THE TECHNICAL SUPPORT SPECIALIST.
IT WAS REPORTED BY THE CUSTOMER THAT THE BD PYXIS¿ MEDSTATION¿ ES Z-EP1 PYXIS MACHINE KEPT GLITCHING. AS SOON AS PULLED SOMETHING OR RECOVERED A STORAGE SPACE IT SAID FATAL ERROR. WHEN A PROVIDER TRIED TO PULL A MEDICATION FOR A PATIENT THE SCREEN FREEZES AND WON¿T DO ANYTHING. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1154241 | BD PYXIS¿ MEDSTATION¿ ES | AUTOMATED DISPENSING CABINET | BRY | CAREFUSION 303, INC. | 323 | 10885403512667 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |