FDA Adverse Event Injury Summary report: N

VANGUARD PS OPEN BOX FEMORAL 75MM RIGHT INTERLOK

MDR report key: 2231809 · Received September 1, 2011

Report

Report Number
1825034-2011-00762
Event Type
Injury
Date Received
September 1, 2011
Date of Event
August 8, 2011
Report Date
August 8, 2011
Manufacturer
BIOMET ORTHOPEDICS
Product Code
JWH
PMA / PMN Number
K023546
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE USER FACILITY IS OUTSIDE OF THE UNITED STATES. NO MEDWATCH REPORT WAS RECEIVED. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER FIFTEEN STATES, "INTEROPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN". THIS IS MDR ONE OF TWO (1825034-2011-00762 THROUGH 00763) FOR THIS EVENT. THIS REPORT SUBMITTED (B)(4), 2011.

Additional Manufacturer Narrative · 1

FEMORAL COMPONENT SHOWED BONE CEMENT EXTENDING INTO THE INTERCONDYLAR BOX. THE SURGICAL TECHNIQUE INSTRUCTS SURGEONS TO REMOVE EXTRUDED CEMENT WITH A CURETTE. THE EXTRUDED BONE CEMENT IMPINGED ON THE POST OF THE TIBIAL BEARING, POSSIBLY LEADING TO THE REPORTED PAIN. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER WARNINGS, NUMBER SEVEN STATES, "COMPLETE PRECLOSURE CLEANING AND REMOVAL OF BONE CEMENT DEBRIS, METALLIC DEBRIS, AND OTHER SURGICAL DEBRIS AT THE IMPLANT SITE IS CRITICAL TO MINIMIZE WEAR OF THE IMPLANT ARTICULAR SURFACES". (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT UNDERWENT TOTAL KNEE ARTHROPLASTY ON (B)(6), 2008. SUBSEQUENTLY, PATIENT WAS REVISED ON (B)(6), 2011, DUE TO PAIN. SURGEON NOTED SOME DISCOLORATION OF TISSUE ADJACENT TO THE IMPLANTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VANGUARD PS OPEN BOX FEMORAL 75MM RIGHT INTERLOK PROSTHESIS, KNEE JWH BIOMET ORTHOPEDICS N/A 609330

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R