FDA Adverse Event Malfunction Summary report: N

GMK FEMORAL CUTTING GUIDE

MDR report key: 2223480 · Received July 6, 2011

Report

Report Number
3005180920-2011-00034
Event Type
Malfunction
Date Received
July 6, 2011
Date of Event
June 21, 2011
Report Date
July 6, 2011
Manufacturer
MEDACTA INTERNATIONAL, SA
Product Code
JWH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A DOCUMENT REVIEW OF THE LOT 096005 ((B)(4)) WAS DONE AND NO ANOMALIES WERE FOUND RELATED TO THE PROBLEM OCCURRED. NO OTHER SIMILAR EVENTS WERE REPORTED CONCERNING THIS LOT. THE BREAKAGE IS THOUGHT TO BE ASSOCIATED TO AN IMPROPER USE DURING PREVIOUS SURGERIES: IN ORDER TO ASSURE A CLOSE CONTACT BETWEEN THE CUTTING GUIDE AND THE DISTAL CUT, HAMMER HITS WERE PROBABLY DONE BY THE SURGEON AND THESE HAVE PROGRESSIVELY WEAKENED THE PIECE LEADING TO ITS BREAKAGE. ON THE BASIS OF MEDACTA'S RISK ANALYSIS, THE FAILURE MODE IS HIGH UNLIKELY TO CAUSE PT HARM SINCE, ALSO IN CASE OF BREAKAGE, THE CUTS COULD BE PERFORMED BECAUSE THE GUIDE IS FIXED TO THE BONE WITH THE PINS, AS HAPPENED IN THIS CASE. SIXTEEN SIMILAR EVENTS WERE REPORTED TO FDA: 2010-00035; 2010-00040; 2010-00041; 2010-00042; 2010-00046; 2010-00047; 2011-00004; 2011-00005; 2011-00006; 2011-00010; 2011-00014; 2011-00016; 2011-00017; 2011-00018; 2011-00021; 2011-00028. AND A FOLLOW-UP TO EXPLAIN THE MODIFICATION/ADJUSTMENT DECIDED, WAS ALREADY SENT ON THE (B)(4) 2011. THE FOLLOW-UP IS APPLICABLE ALSO TO THIS EVENT.

Description of Event or Problem · 1

THE CUTTING GUIDE BROKE INTO TWO PIECES DURING THE SURGERY. NO PT/USER HARM. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GMK FEMORAL CUTTING GUIDE REUSABLE SURGICAL INSTRUMENT FOR KNEE JWH MEDACTA INTERNATIONAL, SA 096005

Patients

Seq Age Sex Outcome Treatment
1 UNK