NEXGEN LPS-FLEX GSF FEMORAL COMPONENT
Report
- Report Number
- 1822565-2011-01877
- Event Type
- Injury
- Date Received
- August 16, 2011
- Date of Event
- September 19, 2009
- Report Date
- July 21, 2011
- Manufacturer
- ZIMMER INC
- Product Code
- JWH
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
EVAL SUMMARY: NO DEVICES OR PHOTOS WERE REC'D; THEREFORE, THE CONDITION OF THE COMPONENTS IS UNK. SURGICAL NOTES WERE NOT PROVIDED. X-RAYS WERE NOT PROVIDED; IT IS UNK WHETHER THE COMPONENTS WERE IMPLANTED WITH THE CORRECT FIT AND ORIENTATION AS PER THE SURGICAL TECHNIQUE. PT FACTORS THAT MAY AFFECT THE PERFORMANCE OF THE COMPONENTS SUCH AS AGE, BONE QUALITY, HEIGHT/WEIGHT, ACTIVITY LEVEL, TYPE OF ACTIVITY (LOW IMPACT VS. HIGH IMPACT) ARE UNK. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED WITH THE INFO PROVIDED. EVAL: REVIEW OF THE DEVICE HISTORY RECORDS DID NOT FIND ANY DEVIATIONS OR ANOMALIES. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADD'L SUBSTANTIVE INFO BE REC'D, THE COMPLAINT WILL BE REOPENED. ZIMMER, (B)(4) CONSIDERS THE INVESTIGATION CLOSED.
IT IS REPORTED THAT THE PT WAS REVISED DUE TO PAIN AND LOOSENING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEXGEN LPS-FLEX GSF FEMORAL COMPONENT | JWH | ZIMMER INC | 60992349 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |