FDA Adverse Event Malfunction Summary report: N

BD PYXIS¿ MEDSTATION¿ ES

MDR report key: 22170156 · Received June 9, 2025

Report

Report Number
2016493-2025-85985
Event Type
Malfunction
Date Received
June 9, 2025
Date of Event
May 15, 2025
Report Date
June 8, 2025
Manufacturer
CAREFUSION 303, INC.
Product Code
BRY
UDI-DI
10885403512667
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHARMACIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE COMPLAINT HISTORY FOR SN (B)(6) WAS PERFORMED IN SALESFORCE WHICH DID NOT LOCATE SIMILAR COMPLAINT(S) WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD FOR SN (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE, 30-JAN-2017 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THE MEDLINK ERROR OCCURRED WHEN TRYING TO PULL MEDICATIONS. A TECHNICAL SUPPORT SPECIALIST HAD DIALED INTO THE STATION AND FOUND THAT THE KPI'S WERE WITHIN THE NORMAL RANGE, NETBIOS WAS ALREADY DISABLED, AND THE MAINTENANCE SERVICE WAS RUNNING. A MEDLINK ERROR OCCURRED WHEN TRYING TO USE THE REMOTE QUEUE WORKFLOW. HOWEVER, THE USER WAS ABLE TO REMOVE MEDS AFTER THE INCIDENT, AND THE ISSUE WAS NOT REPRODUCIBLE. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE TECHNICAL SUPPORT SPECIALIST TROUBLESHOT THE DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED BY THE CUSTOMER THAT WHEN USING THE BD PYXIS¿ MEDSTATION¿ ES, A MEDLINK ERROR OCCURRED WHEN TRYING TO PULL MEDICATIONS. THE CUSTOMER STATED THAT THIS MALFUNCTION CAUSED A DELAY IN DISPENSING MEDICATION TO PATIENTS. HOWEVER, THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1633260 BD PYXIS¿ MEDSTATION¿ ES AUTOMATED DISPENSING CABINET BRY CAREFUSION 303, INC. 323 10885403512667

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown