FDA Adverse Event Malfunction Summary report: N

BD PYXIS¿ MEDSTATION¿ ES AUXILIARY

MDR report key: 21963756 · Received May 5, 2025

Report

Report Number
2016493-2025-78229
Event Type
Malfunction
Date Received
May 5, 2025
Date of Event
April 9, 2025
Report Date
May 5, 2025
Manufacturer
CAREFUSION 303, INC.
Product Code
BRY
UDI-DI
10885403512681
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE COMPLAINT HISTORY FOR SN (B)(6) WAS PERFORMED IN SALESFORCE WHICH DID NOT LOCATE SIMILAR COMPLAINTS WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD FOR SN (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE, 26-OCT-2023 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THE AUXILIARY DRAWER 1, FAILED TO OPEN INTERMITTENTLY. THE FIELD SERVICE ENGINEER ADJUSTED THE CHASSIS RAILS BY LOOSENING THEM A BIT AND POWERED DOWN THE STATION. THEN, THE FIELD SERVICE ENGINEER REMOVED THE DRAWER FROM THE STATION, EXAMINED THE RAILS, FOUND NO DAMAGE. THE DRAWER WAS RETURNED TO THE STATION AND THE CUSTOMER VERIFIED THAT THE DRAWER OPENED NORMALLY AFTER THE RAILS WERE ADJUSTED. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE FIELD SERVICE ENGINEER REPAIRED THE DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN USING THE BD PYXIS¿ MEDSTATION¿ ES AUXILIARY, IT HAD A 6W AUXILIARY DRAWER 1 FH IS CLICKING INSTEAD OF OPENING. THE CUSTOMER REPORTED THERE WAS A DELAY IN DISPENSING MEDICATIONS TO THE PATIENT. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1831645 BD PYXIS¿ MEDSTATION¿ ES AUXILIARY AUTOMATED DISPENSING CABINET BRY CAREFUSION 303, INC. 324 10885403512681

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown