FDA Adverse Event Malfunction Summary report: N

BD PYXIS¿ MEDSTATION¿ ES AUXILIARY

MDR report key: 21960350 · Received May 5, 2025

Report

Report Number
2016493-2025-78075
Event Type
Malfunction
Date Received
May 5, 2025
Date of Event
April 8, 2025
Report Date
May 5, 2025
Manufacturer
CAREFUSION 303, INC.
Product Code
BRY
UDI-DI
10885403512681
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE COMPLAINT HISTORY FOR SN (B)(6) WAS PERFORMED IN SALESFORCE WHICH DID NOT LOCATE SIMILAR COMPLAINTS WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD FOR SN (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE, 13-JUL-2022 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THE NATIONAL TIME PROTOCOL (NTP) SERVER HAD FAILED DUE TO DAYLIGHT SAVING BEING CHECKED. A TECHNICAL SUPPORT SPECIALIST UNCHECKED THE DAYLIGHT SAVING IN THE STATION TEMPORARILY TO RESOLVE. THE HEALTH INFORMATION TECHNOLOGY (HIT) TEAM THEN UPDATED THE NEW NATIONAL TIME PROTOCOL. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE TECHNICAL SUPPORT SPECIALIST AND HIT TEAM TROUBLESHOT THE DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD PYXIS¿ MEDSTATION¿ ES AUXILIARY, THE TIME WAS OFF BY ONE HOUR. THE CUSTOMER REPORTED THAT THE MALFUNCTION RESULTED DELAY IN DISPENSING OF THE MEDICATION TO THE PATIENT. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
458362 BD PYXIS¿ MEDSTATION¿ ES AUXILIARY AUTOMATED DISPENSING CABINET BRY CAREFUSION 303, INC. 324 10885403512681

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown