FDA Adverse Event Malfunction Summary report: N

BD PYXIS¿ MEDSTATION¿ ES

MDR report key: 21691222 · Received March 25, 2025

Report

Report Number
2016493-2025-68660
Event Type
Malfunction
Date Received
March 25, 2025
Date of Event
February 26, 2025
Report Date
March 25, 2025
Manufacturer
CAREFUSION 303, INC.
Product Code
BRY
UDI-DI
10885403512667
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THE DEVICES WERE IN A DISCONNECTED MODE AND OFFLINE ON RSS, LIKELY BECAUSE THEY WERE UTILIZING STATIC IP ADDRESSES. A FIELD SERVICE ENGINEER SWITCHED THE NETWORK SETTINGS FROM STATIC TO DHCP FOR THE ENTIRE FACILITY TO RESOLVE THE ISSUE. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE FIELD SERVICE ENGINEER TROUBLESHOT THE DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN USING THE BD PYXIS¿ MEDSTATION¿ ES, THE DEVICES WERE IN A DISCONNECTED MODE AND OFFLINE ON RSS, WHICH HINDERED USERS FROM ACCESSING MEDICATION FOR PATIENTS. THE CUSTOMER REPORTED THAT THERE WAS A DELAY IN DISPENSING MEDICATIONS TO PATIENTS. THE DELAY WAS NOT CAUSED BY THE PYXIS DEVICE; IT WAS A RESULT OF THEIR DECISION TO SWITCH FROM A STATIC IP TO DHCP, WHICH LED TO A MINOR HOLD-UP IN PROCESSING NEW PATIENTS SINCE THEY HAD TO ENTER THEIR INFORMATION MANUALLY. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1319256 BD PYXIS¿ MEDSTATION¿ ES AUTOMATED DISPENSING CABINET BRY CAREFUSION 303, INC. 323 10885403512667