BD PYXIS¿ MEDSTATION¿ ES
Report
- Report Number
- 2016493-2025-68660
- Event Type
- Malfunction
- Date Received
- March 25, 2025
- Date of Event
- February 26, 2025
- Report Date
- March 25, 2025
- Manufacturer
- CAREFUSION 303, INC.
- Product Code
- BRY
- UDI-DI
- 10885403512667
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THE DEVICES WERE IN A DISCONNECTED MODE AND OFFLINE ON RSS, LIKELY BECAUSE THEY WERE UTILIZING STATIC IP ADDRESSES. A FIELD SERVICE ENGINEER SWITCHED THE NETWORK SETTINGS FROM STATIC TO DHCP FOR THE ENTIRE FACILITY TO RESOLVE THE ISSUE. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE FIELD SERVICE ENGINEER TROUBLESHOT THE DEVICE.
IT WAS REPORTED THAT WHEN USING THE BD PYXIS¿ MEDSTATION¿ ES, THE DEVICES WERE IN A DISCONNECTED MODE AND OFFLINE ON RSS, WHICH HINDERED USERS FROM ACCESSING MEDICATION FOR PATIENTS. THE CUSTOMER REPORTED THAT THERE WAS A DELAY IN DISPENSING MEDICATIONS TO PATIENTS. THE DELAY WAS NOT CAUSED BY THE PYXIS DEVICE; IT WAS A RESULT OF THEIR DECISION TO SWITCH FROM A STATIC IP TO DHCP, WHICH LED TO A MINOR HOLD-UP IN PROCESSING NEW PATIENTS SINCE THEY HAD TO ENTER THEIR INFORMATION MANUALLY. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1319256 | BD PYXIS¿ MEDSTATION¿ ES | AUTOMATED DISPENSING CABINET | BRY | CAREFUSION 303, INC. | 323 | 10885403512667 |