FDA Adverse Event Injury Summary report: N

NEXGEN FLUTED STEMMED TIBIAL COMPONENT

MDR report key: 2162043 · Received July 12, 2011

Report

Report Number
1822565-2011-01623
Event Type
Injury
Date Received
July 12, 2011
Date of Event
February 1, 2011
Report Date
June 15, 2011
Manufacturer
ZIMMER, INC.
Product Code
JWH
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
DC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INFO WAS RECEIVED FROM A DISTRIBUTOR WHO IS NOT REQUIRED TO COMPLETE FORM 3500A. EVAL SUMMARY: SURGICAL NOTES WERE NOT PROVIDED. X-RAYS WERE NOT PROVIDED; IT IS UNK WHETHER THE COMPONENTS WERE IMPLANTED WITH THE CORRECT FIT AND ORIENTATION AS PER THE SURGICAL TECHNIQUE. PT FACTORS THAT MAY AFFECT THE PERFORMANCE OF THE COMPONENTS SUCH AS AGE, BONE QUALITY, HEIGHT/WEIGHT, ACTIVITY LEVEL, TYPE OF ACTIVITY (LOW IMPACT VS HIGH IMPACT) ARE UNK. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED WITH THE INFO PROVIDED. EVALUATION: THE DEVICE HISTORY RECORDS WERE REVIEWED AND FOUND TO BE CONFORMING, INDICATING THE DEVICES WERE MFG TO APPLICABLE SPECIFICATIONS. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL SUBSTANTIVE INFO BE RECEIVED, THE COMPLAINT WILL BE REOPENED. ZIMMER, INC CONSIDERS THE INVESTIGATION CLOSED.

Description of Event or Problem · 1

IT IS REPORTED THAT THE PT WAS REVISED DUE TO LOOSENING OF THE TIBIAL COMPONENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NEXGEN FLUTED STEMMED TIBIAL COMPONENT JWH ZIMMER, INC. 60867920

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention