NEXGEN FLUTED STEMMED TIBIAL COMPONENT
Report
- Report Number
- 1822565-2011-01622
- Event Type
- Injury
- Date Received
- July 12, 2011
- Date of Event
- April 19, 2011
- Report Date
- June 15, 2011
- Manufacturer
- ZIMMER, INC.
- Product Code
- JWH
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DC, US
- Reporter Occupation
- PHYSICIAN
Narratives
INFO WAS RECEIVED FROM A DISTRIBUTOR WHO IS NOT REQUIRED TO COMPLETE FORM 3500A. EVAL SUMMARY: SURGICAL NOTES WERE NOT PROVIDED. SINCE X-RAYS WERE NOT PROVIDED, IT IS UNK WHETHER THE COMPONENTS WERE IMPLANTED WITH THE CORRECT FIT AND ORIENTATION AS PER THE SURGICAL TECHNIQUE. PT FACTORS THAT MAY AFFECT THE PERFORMANCE OF THE COMPONENTS SUCH AS AGE, BONE QUALITY, HEIGHT/WEIGHT, ACTIVITY LEVEL, TYPE OF ACTIVITY (LOW IMPACT VS HIGH IMPACT) ARE UNK. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED WITH THE INFO PROVIDED. EVALUATION: THE DEVICE HISTORY RECORD HAS BEEN REVIEWED AND THE LOT IN QUESTION WAS PRODUCED, INSPECTED AND PACKAGED WITHIN ESTABLISHED AND VALIDATED PROCESS PARAMETERS. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL SUBSTANTIVE INFO BE RECEIVED, THE COMPLAINT WILL BE REOPENED. ZIMMER, INC CONSIDERS THE INVESTIGATION CLOSED.
IT IS REPORTED THAT THE PT WAS REVISED DUE TO LOOSENING OF THE TIBIAL COMPONENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEXGEN FLUTED STEMMED TIBIAL COMPONENT | JWH | ZIMMER, INC. | 60562922 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |