NEXGEN LCCK ART SURFACE W/LOCKING SCREW
Report
- Report Number
- 1822565-2011-01542
- Event Type
- Injury
- Date Received
- June 30, 2011
- Date of Event
- May 30, 2011
- Report Date
- June 3, 2011
- Manufacturer
- ZIMMER, INC.
- Product Code
- JWH
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION SUMMARY: SURGICAL NOTES WERE NOT PROVIDED. X-RAYS WERE NOT PROVIDED; IT IS UNKNOWN WHETHER THE COMPONENTS WERE IMPLANTED WITH THE CORRECT FIT AND ORIENTATION AS PER THE SURGICAL TECHNIQUE. PATIENT FACTORS THAT MAY AFFECT THE PERFORMANCE OF THE COMPONENTS SUCH AS AGE, BONE QUALITY, HEIGHT/WEIGHT, ACTIVITY LEVEL, TYPE OF ACTIVITY ARE UNKNOWN. A DEFINITIVE ROOT CAUSE OF THE SCREW FRACTURE CANNOT BE DETERMINED WITH THE INFORMATION PROVIDED. FURTHERMORE, THE STERILIZATION PROCESS FOR THIS DEVICE WAS VALIDATED IN ACCORDANCE WITH FDA REGULATIONS AND ISO STANDARDS TO A STERILITY ASSURANCE LEVEL (SAL) OF 1.0 X 10 (-6) OR BETTER. THE MANUFACTURING LOT SPECIFIED IN THIS COMPLAINT WAS PROCESSED ACCORDING TO THE VALIDATED STERILIZATION PROCESS PARAMETERS AND MET ALL THE ACCEPTANCE CRITERIA FOR STERILITY RELEASE. THEREFORE, IT IS HIGHLY UNLIKELY THAT THE SPECIFIED DEVICE CAUSED OR CONTRIBUTED TO ANY PATIENT INFECTION. EVALUATION: REVIEW OF THE DEVICE HISTORY RECORDS DID NOT FIND ANY DEVIATIONS OR ANOMALIES. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL SUBSTANTIVE INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED.
IT IS REPORTED THAT WHILE ATTEMPTING TO EXCHANGE THE ARTICULAR SURFACE FOR POSSIBLE INFECTION, THE SURFACE LOCKING SCREW FRACTURED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEXGEN LCCK ART SURFACE W/LOCKING SCREW | JWH | ZIMMER, INC. | 61620936 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Required Intervention |